Term
Turbidity is the measurement (in the lab) |
|
Definition
|
|
Term
When measuring turbidity, color will incrase the reading. |
|
Definition
|
|
Term
When high suspended solids limit the side scatter, the turbidity reading will be higher. |
|
Definition
|
|
Term
Samples filtered through a 0.45 micron filter will not scatter light. |
|
Definition
|
|
Term
Black carbon particles in water will always absorb light |
|
Definition
|
|
Term
Light detected by photocell from suspended solids causes stray light. |
|
Definition
|
|
Term
Stray light is caused by... |
|
Definition
- Glass imperfection
- fingerprints
- scratches
|
|
|
Term
Change in light intensity over time will cause... |
|
Definition
|
|
Term
The advantage of the radio turbidity meter is... |
|
Definition
its ability to color compensate |
|
|
Term
The advantage of the dual beam method of turbidity measurement is... |
|
Definition
the reduction in frequent calibration |
|
|
Term
The type of light required by USEPA in a turbidity meter is ... |
|
Definition
|
|
Term
Volatile Total Solids are dtermined by heating the sample to what temperature? |
|
Definition
|
|
Term
What type of dishes is used in determining total solids? |
|
Definition
- Porcelain
- Silica glass
- platinum
|
|
|
Term
What method of drying is used in the total solids test? |
|
Definition
|
|
Term
What is the minimum weiging accuracy required for the total solids test? |
|
Definition
|
|
Term
What equipment is required to determine the total solids test? |
|
Definition
Pipet, analytical balance, evaporative dish, dessicator, drying oven |
|
|
Term
The purpose of the dessicator is to ... |
|
Definition
Prevent moisture from entering the sample |
|
|
Term
The correct range of residue for the total solids test is ... |
|
Definition
|
|
Term
What percent of samples should be analyzed in duplicate? |
|
Definition
|
|
Term
What reason(s) would there be for weight change during the weighing of the sample? |
|
Definition
- Moisture leaving the sample
- Moisture entering the sample
- sample was warmer than the room temperature
- dessicant needed to be changed
|
|
|
Term
Duplicate samples should be within... |
|
Definition
|
|
Term
Calculate the total solids if the dish weighed 21.1 g and the residue weighed 0.3 g for 100 mL sample. |
|
Definition
|
|
Term
Name an interference to the total solids test |
|
Definition
- oil
- grease
- floating material
- hydroscopic material
|
|
|
Term
The muffle furnace is used to determine the amount of... |
|
Definition
|
|
Term
Calculate the amount of total solids if the dish weighed 20 g and the dish plus residue weighed 21 g with a sample volume of 100 ml. |
|
Definition
|
|
Term
The total solids test will measure... |
|
Definition
- the suspended solids in the sample
- The dissolved solids in the sample
- inorganic solids
- organic solids
|
|
|
Term
According to USEPA specifications, the distance travled by incident light and scattered light within the sample tube shall be... |
|
Definition
|
|
Term
According to USEPA specifications, the angle of light acceptance by detector shall be |
|
Definition
|
|
Term
The spectural peak response shall be between... |
|
Definition
|
|
Term
Dilution water should be filtered through a ... |
|
Definition
|
|
Term
The stock standard for turbidity measurement is... |
|
Definition
|
|
Term
Nephelometric turbidity measurement refers to... |
|
Definition
measuring the light that is reflected into the photocell |
|
|
Term
Manufactured supplied standards equal to the Formazin standard are considered... |
|
Definition
|
|