Term
Define Doppler shift or Doppler frequency |
|
Definition
a change or variation in the frequency of sound as a result of motion between the sound source and the receiver. Greater velocities create greater Doppler shifts. |
|
|
Term
What is it called when source and receiver are approaching each other?
Is the reflected frequency higher or lower than transmitted frequency? |
|
Definition
Positive change or shift
Higher |
|
|
Term
What is it called when source and receiver are moving away from each other?
Is the reflected frequency higher or lower than transmitted frequency? |
|
Definition
Negative change or frequency shift
Lower |
|
|
Term
What does Doppler measure and what are the units? |
|
Definition
Doppler measures frequency shift and the units are Hz |
|
|
Term
What are the typical values of frequency shift in sonography?
|
|
Definition
20 Hz - 20 kHz, audible.
We still use 2 MHz to 10 MHz transducers to perform a Doppler ultrasound study, but the Doppler shift (which is a change) ranges from 20-20,000 Hz. |
|
|
Term
Frequency shift is always related to __________. |
|
Definition
|
|
Term
The Doppler shift is a low frequency that "rides" or is "carried" on top of the much higher transducer frequency. The process of extracting the low Doppler frequency is called __________. |
|
Definition
|
|
Term
Give the Doppler Equation and what does each part stand for? |
|
Definition
Δf=2foV(cosθ)/c
Δf = Doppler shift (kHz)
fo = Transmitted frequency (MHz)
V = Velocity of blood flow (cm/s)
cosθ = cosine angle between US and blood flow direction
c = speed of sound in soft tissue (1540 m/s) |
|
|
Term
Doppler shift is __________ related to the blood cell speed frequency of the transducer cosine of the angle between flow and the sound beam. |
|
Definition
|
|
Term
Doppler shift is __________ related to the speed of sound in the medium. |
|
Definition
|
|
Term
Doppler measures __________ not _________; therefore, in order to accurately determine the __________, the angle between the directions of flow and sound beam must be known. |
|
Definition
|
|
Term
What is the difference between "speed" and "velocity"? |
|
Definition
Speed is magnitude only, whereas, velocity is magnitude (speed) and direction. |
|
|
Term
Doppler frequency depends on __________. The magnitude of shift depends upon the __________ __ ___ __________ between the sound beam and direction of the motion. |
|
Definition
direction
cosine of the angle |
|
|
Term
What is the cosine for 0°? |
|
Definition
|
|
Term
What is the cosine for 60°? |
|
Definition
|
|
Term
What is the cosine for 90°? |
|
Definition
|
|
Term
What does the 2 in the Doppler equation represent? |
|
Definition
In clinical Doppler, there is a double Doppler shift. The first occurs when the sound strikes the cell. The second shift results from the moving blood cell reflecting the wave back to the transducer. |
|
|
Term
If the sound beam and flow direction are not parallel, what determines how much of the velocity is measured? |
|
Definition
The percentage of the true velocity that is measured depends on the cosine of the angle between the sound beam and the direction of motion. |
|
|
Term
What is the relationship between the cosine θ and the Doppler shift? |
|
Definition
Mathematically, the Doppler shift is directly related to the cosine of the angle between the direction of flow and the direction of sound. If the cosine θ is reduced by half, the measured Doppler shift will be halved. |
|
|
Term
What is the relationship between the transmitted frequency and the Doppler shift? |
|
Definition
Mathematically, the Doppler shift is directly related to the frequency of the transmitted sound. In other words, if the transducer's frequency is doubled, the measure Doppler shift will also be doubled. |
|
|
Term
What is the relation between actual velocity and measured velocity when the blood moves parallel to the sound beam? |
|
Definition
When the angle between the direction of motion and sound is 0° or 180°, flow is parallel to the sound beam. The measured and true velocities are equal. Cosine θ is 1, indicating flow toward the transducer, whereas cosine 180° is -1, indicating flow away from the transducer. |
|
|
Term
What is the relationship between actual and measured velocities when blood moves perpendicular to the sound beam? |
|
Definition
When the direction of flow is perpendicular (90°) to the sound beam, the measured velocity is zero because the cosine of 90° is zero. Thus, Doppler shifts and velocities cannot be measured with perpendicular incidence. |
|
|
Term
What is the relationship between actual velocity and measured velocity when the blood moves at a 60° angle to the beam? |
|
Definition
Since the cosine of 60° is 0.5, the measured velocity at 60° is one-half the actual velocity. When blood travels at 2 m/s at a 60° angle to the sound beam, Doppler reports the velocity as 1 m/s. |
|
|
Term
How many crystals are required for CW Doppler and why? |
|
Definition
Two crystals in the transducer:
one crystal is constantly transmitting
the other is continuously receiving |
|
|
Term
What is the greatest advantage of CW Doppler? |
|
Definition
Able to measure very high velocities accureately. |
|
|
Term
What is the prime disadvantage for CW Doppler? |
|
Definition
Because echoes arise from the entire length of overlap between the transmit and receive beams, there is range (depth) ambiguity. |
|
|
Term
How many crystals doe PW Doppler use and why? |
|
Definition
One crystal, alternates between sending and receiving. |
|
|
Term
What is the greatest advantage of PW Doppler? |
|
Definition
Echoes arise only from the area of interrogation, the sample volume or gate, We adjust the receive gate. This is called a range (depth) resolution or range specificity or freedom from range ambiguity artifact. |
|
|
Term
What is the primary disadvantage of PW Doppler? |
|
Definition
Inaccurate measurement of high velocity signals. This inaccuracy is called aliasing, a false identity. |
|
|
Term
What is the term for a study that simultaneously perform imaging and PW Doppler interrogations? |
|
Definition
|
|
Term
What is the X-axis of a Doppler spectrum represent? |
|
Definition
|
|
Term
What is the Y-axis of a Doppler spectrum represent? |
|
Definition
Doppler shift or velocity |
|
|
Term
|
Definition
Aliasing is the most common error associated with Doppler ultrasound. An alias is a false identity. The false identity in PW Doppler is that very high velocities in one direction are incorrectly displayed as going in the opposite direction.
Flow above the baseline is going toward the transducer, whereas flow below the baseline is going away from the transducer. With aliasing, the high velocity flow reaches the very top of the spectral display then wraps around and reappears at the bottom. |
|
|
Term
What is the very top of the spectral display called? |
|
Definition
|
|
Term
What is the Nyquist frequency or limit? |
|
Definition
The Doppler frequency at which aliasing occurs, equal to 1/2 the PRF. |
|
|
Term
What is the equation for the Nyquist limit? |
|
Definition
|
|
Term
When aliasing occurs, what what 2 tactics could be utilized to eliminate aliasing? |
|
Definition
Either:
- reduce the Doppler shift (reduce the speed)
- raise the Nyquist limit (raise the speed limit)
|
|
|
Term
What conditions create aliasing? |
|
Definition
- aliasing occurs only with pulsed Doppler
- aliasing occurs when the Doppler sampling rate is too low in comparison to the measured blood velocities
|
|
|
Term
How does transducer frequency affect aliasing? |
|
Definition
- Higher frequency transducers create more aliasing.
- Doppler shift (Δf) is (=) directly related to the transducer's frequency (fo). The Doppler shift is percentage of the transducer frequency.
- At a particular velocity, higher frequency transducers create higher Doppler shifts and lower frequency transducers create lower Doppler shifts.
- Smaller Doppler shifts from lower frequency transducers are less likely to exceed the Nyquist limit. Therefore, aliasing artifacts are less common with lower frequency transducers.
|
|
|
Term
|
Definition
Color Doppler is a form of 2-D, multigated Doppler. Color Doppler is pulsed ultrasound so it is subject to range resolution or specificity and aliasing. Color Doppler measures mean blood flow velocities and identifies function. |
|
|
Term
What do the colors in a velocity mode map mean? |
|
Definition
- The colors provide information on flow direction and velocity.
- Colors in the top half represent blood cells moving toward the transducer, while colors in the bottom half represent away from the transducer.
|
|
|
Term
What do the colors in a variance mode map mean? |
|
Definition
- Similar to velocity mode maps, the colors in a variance mode map indicate flow direction. Colors in the top half represent blood cells moving toward the transducer, while colors in the bottom half represent away from the transducer.
- In addition, the variance maps display different colors side to side. Colors on the left side of the map indicate laminar flow.
- Colors on the right side of the map indicate turbulent flow.
- Remember L--left, laminar
|
|
|
Term
What are Doppler Packets? |
|
Definition
With color Doppler, multiple ultrasound pulses are used to accurately determine blood velocities. These multiple pulses are called a packet, or ensemble. |
|
|
Term
What are the advantages of Doppler packets? |
|
Definition
Packets composed of a larger number of pulses (larger packets or longer ensemble lengths) have two advantages:
- more accurate velocity measurement
- increased sensitivity to low flow
|
|
|
Term
What are the disadvantages of Doppler packets? |
|
Definition
Packets with more pulses have these disadvantages:
- more time needed to acquire data
- reduced frame rate
- decreased temporal resolution
|
|
|
Term
What is Power Doppler and what are some common synonyms? |
|
Definition
Power mode only identifies the presence of a Doppler shift. It does not evaluate speed or direction. Power mode is non-directional color Doppler.
Power Doppler is also called energy mode or color angio. |
|
|
Term
What are the advantages of power Doppler? |
|
Definition
- increased sensitivity to low flow or velocity, such as venous flow or flow in small vessels
- unaffected by Doppler angles, unless the angle is exactly 90°
- no aliasing, since the velocity information is ignored
|
|
|
Term
What are the disadvantages of power Doppler? |
|
Definition
- no measurement of velocity or direction
- lower frame rates (reduced temporal resolution) than conventional color flow Doppler
- susceptible to motion of the transducer, patient, or soft tissues, which may result in a burst of color, or flash artifact
|
|
|
Term
What is ghosting or clutter? |
|
Definition
- Doppler shifts generally arise from moving blood. However, slowly moving anatomy, such as heart muscle or pulsating vessel walls, may create very low frequency Doppler shifts.
- On spectral displays, these low frequency Doppler shift artifacts are called clutter.
- With color Doppler, they are called ghosting artifact.
|
|
|
Term
How is clutter or ghosting artifact eliminated? |
|
Definition
- A wall filter is used to eliminate low frequency Doppler shifts from moving anatomy rather than from moving blood cells. In essence, wall filters serve as a "reject" for Doppler.
- Wall filters are also called, high pass filter
|
|
|
Term
|
Definition
- Crosstalk is a special form of "mirror image" artifact, arised only with spectral Doppler.
- Crosstalk appears as an identical Doppler spectrum both above and below the baseline. It appears as bidirectional flow when in reality it is unidirectional.
|
|
|
Term
What generally causes crosstalk? |
|
Definition
- Doppler gain is set too high (electronic crosstalk)
- the incident angle is near 90° between the sound beam and the flow direction, when flow is at the beam's focus.
|
|
|
Term
What is spectral analysis? |
|
Definition
- Performed to extract or identify the individual frequencies making up the complex signal. It is used to interpret individual velocities in the signal.
|
|
|
Term
How many many methods of spectral analysis is currently used, what are they called, and when are they used? |
|
Definition
Two methods of spectral analysis are currently used:
- For PW or CW Doppler: Fast Fourier Transform (FFT)
- For color flow Doppler: autocorrelation or correlation fuction
|
|
|
Term
What is Fast Fourier Transform (FFT) and what are the advantages? |
|
Definition
A digital (computer-based) technique, is used to process both PW and CW Doppler signals. FFT have two advantages
- exceedingly accurate
- displays all individual velocity components that make up the complex reflected signal
|
|
|
Term
What is the significance if displaying all component frequencies with FFT? |
|
Definition
The spectral display of an FFT distinguishes laminar flow from turbulent flow. |
|
|
Term
|
Definition
The digital (computer-based) technique used to analyze color flow Doppler |
|
|
Term
What is Normal Incidence? |
|
Definition
Normal incidence means that the incident sound beam strike the boundary at exactly 90° |
|
|
Term
[image]
In color Doppler image A, why does the vessel appear without color? |
|
Definition
In this case, the angle between the direction of flow and the direction of the sound beam is 90°. Since the cosine of 90° is zero, no Doppler shift is created or measured, and no color appears in the vessel. |
|
|
Term
[image]What will make color appear in vessel A? |
|
Definition
For color to appear in the vessel, angle other than 90° must be created between the direction of flow and the direction of the sound beam. |
|
|
Term
[image]
Why does vessels B and C have color and A does not?
Why is vessel B blue and C red? |
|
Definition
- In vessel A, the angle of incidence is 90° which makes the cosine θ zero so there will be no Doppler shift.
- In vessel B, the color box has been steered to the left, creating an oblique angle (>90°); thus, the cosine θ is negative. Therefore, the flow is away from the transducer's beam.
- In vessel C, the color box has been steered to the right, creating an acute angle (<90°); thus, the cosine θ is positive. Therefore, the flow is towards the transducer's beam.
|
|
|
Term
[image]
In color Doppler image A, why is there no color within the vessel? |
|
Definition
Even though the vessel is at an angle, the color box strikes the vessel at 90°; thus, the cosine θ is zero. |
|
|
Term
[image]
What will make color appear in vessel A? |
|
Definition
For color to appear in the vessel, an angle other than 90° must be created between the direction of flow and the direction of the sound beam. In image B, the color box is steered to the right, creating a shift. In image C, the color box is steer more to the right, creating a closer angle to zero, where the cosine θ is 1 |
|
|
Term
[image]
What single control on an ultrasound system can make color appear throughout the box as shown in image B? |
|
Definition
Image B was created by setting the color gain too a level that is too high.
When the color gain is set to an inappropriately high level, color appears throughout the color box. A term that effectively describes this is color confetti. |
|
|
Term
[image]What single control on the ultrasound system can make color disappear as shown in image C? |
|
Definition
The color gain is set too low in image C that all the color has disappeared. |
|
|
Term
[image]What single control on the ultrasound system can create image B? |
|
Definition
The color gain is set too high in image B |
|
|
Term
[image]What single control on an ultrasound system can make color disappear throughout the color sector? |
|
Definition
Image C was created with too low a color gain setting. |
|
|
Term
[image]What alteration was responsible for the change in the appearance of the spectrum from image A to image B? |
|
Definition
Image A is high quality. Image B was created by increasing the pulsed Doppler gain to a level that is too high. Gray scale noise appears throughout the spectrum when the pulsed Doppler gain is set too high. |
|
|
Term
[image]What alteration was responsible for the change in the appearance of the spectrum from image A to image C? |
|
Definition
Image C was created by decreasing the pulsed Doppler gain to a level that is too low. When the pulsed Doppler gain is set to an inappropriately low level, all gray scale will disappear in the image. Not only does the noise vanish, so does the meaningful Doppler spectrum. |
|
|
Term
[image]What artifact appears in image A? |
|
Definition
Aliasing is present in the color Doppler image in image A.
Image B is a zoomed image of the flow in the artery (upper vessel). The central flow stream is blue, indicating flow toward the transducer, or from right to left (blue arrow).
The flow streams closer to the vessel walls are red, indicating flow away from the transducer, or from left to right (red arrow). It is unlikely that there is flow in different directions within a single vessel. Aliasing should be considered as a possible source of this peculiar color variation, whenever the color changes directly from red to blue or vice versa. |
|
|
Term
[image]Is the blue area in this image aliasing or flow reversal? |
|
Definition
To answer the question, look at the colors that are adjacent to the blue area. The transition of colors is from red to black to blue (yellow arrow). On the color map, red to black to blue forces us to move through the center of the map. This color transition indicates flow reversal.
Aliasing on the image would appear as green adjacent to yellow. |
|
|
Term
[image]What adjustment did the songrapher perform to create image B? |
|
Definition
Increased the wall filter to a very high level. Two changes appear when comparing images A and B. First, the color in the vein has vanished. Second, the ghosting artifact along the deeper wall of the artery (yellow arrow, image A) has also disappeared. Color in the artery is limited to the lumen and no longer extends deeper than the vessel wall. |
|
|
Term
[image]What adjustment did the sonographer perform to change the upper spectrum (image A) into the lower spectrum (image B). |
|
Definition
Increased the wall filter to a higher level. Wall filters are used in both color flow and spectral Doppler. |
|
|
Term
[image]What adjustment did the sonographer perform to create the image B? |
|
Definition
The sonographer increased the wall filter to a higher level. The color in the vein has vanished. The wall filter selectively eliminates low frequency Doppler shifts. |
|
|
Term
[image]What adjustment did the sonographer perform to create image B? |
|
Definition
The sonographer increased the scale. In image A, aliasing appears in the artery. The appearance of adjacent colors, red, yellow, light blue, and dark blue, is evidence of aliasing. In image B, aliasing is no longer present in the artery. This is consistent with increasing the scale. |
|
|