Term
Functional Saturation
(SaO2) |
|
Definition
|
|
Term
Oxyhemoglobin Fraction
(O2Hb) |
|
Definition
O2Hb/ (O2Hb + RHb + COHb + MetHb) x 100 |
|
|
Term
|
Definition
(1.37 x Hb x O2Hb%/100) + (0.003 x PaO2)
Measures O2 content per 100mLs of blood |
|
|
Term
Normal Oxygen Content
(CaO2) |
|
Definition
|
|
Term
|
Definition
|
|
Term
|
Definition
|
|
Term
|
Definition
O2 sats become independent of PaO2 |
|
|
Term
|
Definition
(CaO2 x 10) x CO
Converts deciliters to liters |
|
|
Term
|
Definition
O2 delivered - O2 returned
([CaO2 - CvO2) x 10]xCO
[1.37 x (Hb x O2sat/100- O2venous/100 x Hb)] x 10 x CO |
|
|
Term
|
Definition
VO2=[1.37 x (15 x100/100- 75/100 x 15)] x 10 x 5.0
VO2= 250 cc/min |
|
|
Term
|
Definition
Makes use of the measurement of light absorbance to determine the concentration of solute in clear solutions (Beer-Lambert)
Estimate arterial hemoglobin saturation by measuring the transmission of two wavelengths of light through a pulsatile vascular bed
Distinguish light absorbance of arterial blood from other tissues by looking for blood pulsations |
|
|
Term
|
Definition
Based on the assumption that arterial blood is the only pulsatile absorpber
By using two wavelengths at 660 nm and 940 nm as well as pulsatile flow, the oximeter is able to determine the concentrations of O2Hb and RHb
These are used to form a ratio (R) |
|
|
Term
|
Definition
Directly related to hemoglobin saturation
From experimental data, R is plotted to give corresponding O2 saturation |
|
|
Term
|
Definition
Dyes
COHb and MetHb
Wavelength Uncertainty
Signal to Noise Ratio |
|
|
Term
|
Definition
If any light absorbing species other than RHb and O2Hb are present, the oximeter cannot accurately estimate saturation
Methylene Blue and Indigo Carmine |
|
|
Term
|
Definition
Can produce sudden, large decreases in SpO2 |
|
|
Term
Indigo Carmine and Indocyanine Green |
|
Definition
|
|
Term
|
Definition
Both absorb light at one or both used wavelengths |
|
|
Term
|
Definition
Appears to the oximeter as O2Hb and will seriously over-estimate arterial saturation |
|
|
Term
|
Definition
Absorbs both wave lengths equally
As both numerator and denominator are increased, R approaches 1.0
From the algorithm, if R=1 then the satration will be read as 85% |
|
|
Term
|
Definition
LED's may emit a non-monochromatic light
This is dealt with by having multiple algorithms in the software
The connector for the sensor cable is coded for the appropraite algorithm |
|
|
Term
|
Definition
The detector may respond to room light
Weak pulses can make the pulsatile component extremely small and difficult to separate from tissue absorption
The Penumbra effect |
|
|
Term
|
Definition
If the detector is improperly placed so the light passes through the fingertip at a grazing incidence, a correct heart rate may be displayed falsely low saturation
May be a similar effect to R=1.0 |
|
|
Term
|
Definition
B/C of varying gain necessary in a variety of clinical situations changes in the signal do not correlate well with peripheral perfusion
Patient movement can cause what the oximeter interprets as pulsations |
|
|
Term
|
Definition
Tricuspid Insufficiency
Gravitational effect |
|
|
Term
|
Definition
Electromagnetic radiation can be picked up by the sensor cable |
|
|
Term
|
Definition
Most claim +-2% x one standard deviation for values 70-100%
Uncertainity increases to +-3 with saturation values of 50-70%
For saturations above 70% the saturation should be within 2% of the actual saturation 68% of the time
Within 4% 95% of the time
|
|
|
Term
|
Definition
Lung to finger circulation times are longer than lung to ear times
As a result response times are faster with ear probes |
|
|
Term
|
Definition
Can increase response time |
|
|