Term
The principal effect of tissue damage from varying magnetic and radiofrequency appears to be related to the
a. Physical alteration of the subatomic particles in the nucleus
b. Deposition of heat in the tissues exposed to these fields
c. Chemical alteration of the metabolic compounds of the cells
d. Direct ionization of orbital electrons of the atoms in the cells. The structure of interest extends outside of the imaging volume |
|
Definition
b. Deposition of heat in the tissues exposed to these fields |
|
|
Term
During a magnetic resonance imaging procedure, all of the following metallic implants are at risk for being dislodged except:
a. Intracranial aneurysm clips
b. Prosthetic heart valves
c. Auditory implants
d. Cholecystectomy clips |
|
Definition
|
|
Term
In order to avoid surface burns of the patient during a magnetic resonance imaging procedure, it is important to:
a. Place EKG leads as close to the skin as possible
b. Coil all leads into loops and place them under the patient's gown or clothing
c. Place additional insulation around all lead wires
d. Coat all lead wires with a conducting solution of paste |
|
Definition
c. Place additional insulation around all lead wires |
|
|
Term
Earplugs or non-magnetic headphones are commonly used during a magnetic resonance imaging procedure to:
a. Prevent temporary or permanent hearing damage in the patients
b. Protect the external auditory canals of the patients from mechanical injuries
c. Prevent the electrical burns to the tympanic membranes of the patients
d. Reduce the production of eddy currents in the cochlea of the patients
|
|
Definition
a. Prevent temporary or permanent hearing damage in the patients |
|
|
Term
The most common reason for the inability of a patient to successfully complete a magnetic resonance imaging procedure is:
a. Magnectic induced cardiac irregularities
b. Foreign objects in the eye
c. Claustrophobia
d. Intolerance to loud sounds |
|
Definition
|
|
Term
A rare but potential hazard with superconducting magnetic systems is the release of helium or nitrogen into the imaging room resulting in the:
a. Rapid drop of temperature or hyperthermia of the patient
b. Reduction of oxygen levels of the atmosphere and the asphyxiation of the patient
c. Increased conductivity of the atmosphere and the possibility of electrical shock
d. Increased combustibility of the atmosphere and the possibility of an explosion |
|
Definition
b. Reduction of oxygen levels of the atmosphere and the asphyxiation of the patient |
|
|
Term
In most modern 1-2 tesla units a fringe field level of less than 5 gauss will require a distance of at least:
a. 3-5 feet from the bore of the magnet
b. 10-15 feet from the bore of the magnet
c. 20-25 or more feet from the bore of the magnet
d. 50-75 or more feet from the bore of the magnet |
|
Definition
c. 20-25 or more feet from the bore of the magnet |
|
|
Term
When a magnetic resonance unit has a large fringe field, the movement of large metallic objects such as cars or elevators may:
a. Cause ripples in the magnetic field that can hinder image quality
b. Increase the heating in electrical wires resulting in voltage brownouts
c. Cause alterations in the recording media used to store the images
d. Increase the strength of the magnetic field above its safety limits |
|
Definition
a. Cause ripples in the magnetic field that can hinder image quality |
|
|
Term
Magnetophosphene or the "seeing of stars" is a reversible phenomenon occurring in the eyes caused by an individual's exposure to strong
1. Static magnetic fields
2. Time-varying magnetic fields
3. Radiofrequencies
a. 1 only
b. 2 only
c .3 only
d. 1, 2, & 3 |
|
Definition
|
|
Term
What is the main biologic hazard if a surface coil for magnetic resonance imaging of the knee is inadvertently angled with respect to the circumferential coil?
a. The patient may experience erythema due to the production of ultraviolet radiation by the device
b. The patient may experience skin burns due to the excessive heating of the device
c. The patient may experience frostbite due to the excessive cooling of the coil
d. The patient may experience mechanical trauma due to the movement of the device within the field
|
|
Definition
b. The patient may experience skin burns due to the excessive heating of the device |
|
|
Term
In order to reduce the potential for biological effects during magnetic resonance imaging studies, the Food and Drug Administration has limited the maximum strength of the static magnetic field to no more than 2 tesla. This recommendation has been made to reduce the:
a. Potential of high voltage shocks
b. Fringe field limits to less than 100 feet
c. Potential for unwanted biologic effects
d. Potential of radiofrequency interference in adjacent electrical devices |
|
Definition
c. Potential for unwanted biologic effects |
|
|
Term
Magneto hemodynamics is best defined as the interactions that can occur:
a. In cells, organelles, and macromolecules exposed to strong magnetic fields
b. In structural components and building materials when exposed to strong magnetic fields
c. Electrical recording systems when exposed to RF fields
d. In electromagnetic radiation when exposed to strong magnetic fields |
|
Definition
a. In cells, organelles, and macromolecules exposed to strong magnetic fields |
|
|
Term
One of the concerns during a magnetic resonance imaging study is the failure to provide adequate cooling of the patient during the study. In order to prevent discomfort to the patient from heating caused during the production of radiofrequencies and other ambient conditions, the imaging room should provide for all the following EXCEPT:
a. A high humidity
b. A low humidity
c. A high air flow rate
d. A low room temperature |
|
Definition
|
|
Term
The greatest mechanical risk associated with the operation of a modern superconducting magnet is the danger associated with the:
a. Induced electrical currents in surface coils due to varying radiofrequencies
b. Rapid cooling of the unit due to venting of supercooled liquid gases
c. Rapid acceleration ferromagnetic projectile
d. Reversal of magnetic polarity and the effects of sudden molecular realignments
|
|
Definition
c. Rapid acceleration ferromagnetic projectiles |
|
|
Term
A reduction of the blood flow in the vessels due to the presence of electrolytes in a strong static magnetic field is related to the formation of:
a. Ferro polar cells
b. Vasodilation chemicals
c. Coagulation elements
d. Eddy currents |
|
Definition
|
|
Term
A reduction of the blood flow in the vessels due to the presence of electrolytes in a strong static magnetic field is related to the formation of:
a. Ferro polar cells
b. Vasodilation chemicals
c. Coagulation elements
d. Eddy currents |
|
Definition
|
|
Term
The greatest concern for a patient with a pacemaker undergoing a magnetic resonance imaging procedure appears to be the potential for the:
a. Physical displacement of the leads
b. Suppression of the pacing activity
c. Damage to the power supply
d. Acceleration of the pacing rate |
|
Definition
b. Suppression of the pacing activity |
|
|
Term
Research indicates that at usual magnetic resonance imaging levels, the fetus appears to be:
a. Highly sensitive to static magnetism fields
b. Highly sensitive to magnetic resonance radiofrequencies
c. Highly sensitive to time-varying magnetic fields
d. Relatively insensitive to magnetic resonance related stimuli |
|
Definition
d. Relatively insensitive to magnetic resonance related stimuli |
|
|
Term
Which of the following conditions is most often associated with the delayed clearance of gadoteridol
(ProHance) from the body?
a. Renal insufficiency
b. Cardiac insufficiency
c. Gastritis
d. Cholecystitis |
|
Definition
|
|
Term
The main route by which the human immunodeficiency virus (HIV) enters the body is through the:
a. Respiratory tract
b. Circulatory system
c. Urinary tract
d. Digestive system |
|
Definition
|
|
Term
The muscles groups that should be employed for lifting heavy objects are located in the:
1. Arms
2. Legs
3. Back
a. 1 & 2 only
b. 1 & 3 only
c. 2 & 3 only
d. 1, 2, & 3 |
|
Definition
|
|
Term
The two most common methods for the control of the spreading of contagious diseases are:
a. vaccination and inoculation
b. Decompression and contamination
c. Disinfection and isolation
d. Inoculation and asepsis
|
|
Definition
c. Disinfection and isolation |
|
|
Term
Hepatitis B virus is frequently transmitted from a patient to a healthcare worker through:
a. The improper handling of urine
b. Direct contact with the patient's skin
c. Droplets caused by sneezing or coughing
d. Accidental needle stick injuries |
|
Definition
d. Accidental needle stick injuries |
|
|
Term
The common means by which infections are spread is by:
a. Airborne particulates
b. Indirect (fomite) contact
c. Direct(touching) contact
d. Endogenous contact |
|
Definition
c. Direct(touching) contact |
|
|
Term
The failure to keep an IV bottle above the level of the patient during an IV infusion is most likely to result in:
a. Infiltration of the IV solutions into the surround tissues
b. Bacterial contamination of the IV fluids
c. Blood clot formation in the needle or line
d. An increase in the infusion rate of the contrast agent |
|
Definition
c. Blood clot formation in the needle or line |
|
|
Term
When a patient is connected to the chest tube drainage equipment, it is important to_______when moving the patient.
1.Clamp off the chest tube
2. Raise the bottle above the chest
3. Avoid kinks in the drainage tube
a. 1 only
b. 2 only
c. 3 only
d. 1,2, & 3 |
|
Definition
|
|
Term
A tracheostomy is a surgical procedure which is performed to:
a. Reduce the pressure which can buildup in the heart
b. Help relieve a severe blockage of the upper respiratory tract
c. Drain off excessive fluids which have collected in the pleura
d. Sample or portion a lesion for laboratory testing |
|
Definition
b. Help relieve a severe blockage of the upper respiratory tract |
|
|
Term
Syncope or fainting is a mild form of neurogenic shock that may occur:
a. After a highly unpleasant or painful event
b. Following a sudden change of posture which increases the blood pressure
c. After the exposure to an air-borne contamination
d. Following exposure to a pathogen |
|
Definition
a. After a highly unpleasant or painful event |
|
|
Term
The signs of cardiac arrest will normally include:
1. Apnea
2. Absence of the pulse
3. Falling blood pressure reading
a. 1 only
b. 2 only
c. 3 only
d. 1, 2, & 3 |
|
Definition
|
|
Term
The human brain can survive without oxygen for about:
a. 2-4 minutes
b. 7-9 minutes
c. 11-14 minutes
d. 15-20 minutes |
|
Definition
|
|
Term
The cycle of infection can be broken by:
1. Destruction of the infectious organism
2. Removal of the source of infection
3. Preventing the means by which an infection is spread
a. 1 & 2 only
b. 1 & 3 only
c. 2 & 3 only
d. 1, 2, & 3 |
|
Definition
|
|
Term
All of the following are appropriate indications for the ordering of a MR imaging procedure of the head EXCEPT:
a. The evaluation of cerebral infarcts
b. The confirmation of brain death
c. The assessment of metastatic disease
d. The changes associated with multiple sclerosis |
|
Definition
a. The evaluation of cerebral infarcts |
|
|
Term
During vascular imaging of the head which of the following imaging technique is likely to be most sensitive to for fast flow -related vascular lesions such as small aneurysm partial vascular occlusions?
a. Three-dimensional time of flight
b. Two-dimensional time of flight
c. Three-dimensional phase contrast
d. Multi-dimensional phase contrast |
|
Definition
a. Three-dimensional time of flight |
|
|
Term
During the acquisition of axial MR images of the cervical spine, a saturation pulse placed anterior to the cervical spine
may help to reduce:
a. Motion artifacts related to swallowing
b. Flow artifacts related to carotid arterial blood flow
c. Flow artifacts related to the movement of the cerebral spinal fluid
d. Motion artifacts to the movement of air through the bronchial tree |
|
Definition
a. Motion artifacts related to swallowing |
|
|
Term
During MR imaging of the heart and mediastinum which of the following is normally used to reduce motion artifacts?
a. Flow compensators and multiple saturation pulses
b. Thoracic compression and blood flow suppression devices
c. Phase-encoding and motion suppression programs
d. Respiratory compensation bellows and an EKG gated technique |
|
Definition
d. Respiratory compensation bellows and an EKG gated technique |
|
|
Term
During MR imaging of the pelvis, all of the following are used to improve the quality of images EXCEPT:
a. The placement of padding posterior to the patients knees
b. Instruction the patient to breath form the chest and upper abdomen
c. The placement of the arms of the patient over the mid-pelvis
d. The use of peripheral gating techniques |
|
Definition
c. The placement of the arms of the patient over the mid-pelvis |
|
|
Term
A magnetic resonance image of the lumbar spine is obtained in the sagittal plane using a T2 weighted image sequence. Which of the following structures are most likely to produce the highest signal intensity?
a. A desiccated intervertebral disk
b. The body of the vertebra
c. The nucleus pulposus of a normal disk
d. The abdominal aorta or inferior vena cava |
|
Definition
c. The nucleus pulposus of a normal disk |
|
|
Term
Which of the following imaging planes is normally used to demonstrate the relationship of the inferior vena cava to the diaphragm and heart
a. Sagittal image
b. Transverse image
c. Coronal image
d. Oblique image |
|
Definition
|
|
Term
[image]
This magnetic resonance image is best described as:
a. An axial image of the knee
b. A coronal image of the knee
c A sagittal image of the knee
d. An oblique image of the knee |
|
Definition
b. A coronal image of the knee |
|
|
Term
[image]
The appearance of the knee image is most consistent with an image obtained using a:
a. T1 weighted imaging sequence
b. T2 weighted imaging sequence
c. Proton-density imaging sequence
d. subtraction imaging sequence |
|
Definition
b. T2 weighted imaging sequence |
|
|
Term
[image]
The structure represented by number 10 corresponds to the:
a. Medial condyle of the femur
b. Intercondylar eminence
c. Posterior cruciate ligament
d. Medial meniscus |
|
Definition
|
|
Term
[image]
The structure represented by number 1 corresponds to the:
a. Lateral intercondylar spine
b. Anterior cruciate ligament
c. Vastus lateralis muscle
d. Lateral meniscus |
|
Definition
c. Vastus lateralis muscle |
|
|
Term
[image]
The insertion of the posterior cruciate ligament to the distal femur is demonstrated at the arrow labeled:
a. Number 3
b. Number 4
c. Number 6
d. Number 8 |
|
Definition
|
|
Term
Which of the following techniques is often applied during a magnetic resonance imaging procedure of the heart to evaluate flow-related
artifacts in the blood vessels?
a. A subtraction technique
b. A multi-filtered technique
c. A gated technique
d. A rapid acquisition technique |
|
Definition
|
|
Term
A high intraluminal signal from within the jugular vein or dorsal sinus is most often indicative of a/an:
a. Increase blood flow value
b. Obstruction defect in the vessel
c. Reduction in the luminal thickness of the vessel
d. Varicose vessel |
|
Definition
b. Obstruction defect in the vessel |
|
|
Term
A magnetic resonance image of the thoracic spine is obtained using a T1 weighted imaging sequence. All of the following are normally associated with the production of a low-intensity signal EXCEPT:
a. The cortical bone of the vertebral body
b. The intervertebral disks
c. The cancellous bone of the vertebral body
d. The spinal cord |
|
Definition
|
|
Term
The use of paramagnetic contrast agents in conjunction with a magnetic resonance imaging study is most beneficial for the enhancement of:
a. The structures of the central nervous system
b. Spinal cord compression resulting from disk abnormalities
c. Abdominal cysts and scars
d. Structural abnormalities of the bone or bone-like tissues |
|
Definition
a. The structures of the central nervous system |
|
|
Term
[image]
This magnetic resonance image of the left shoulder is best described as a:
a. T1 weighted image
b. T2 weighted image
c. Fat suppression image
d. Contrast-enhanced image |
|
Definition
|
|
Term
[image]
This magnetic resonance image of the shoulder was obtained in:
a. A coronal plane
b. An axial plane
c. A sagittal plane
d. Off-axial plane |
|
Definition
|
|
Term
[image]
The structure represented by number 3 corresponds to the:
a. Trapezius muscle
b. Deltoid muscle
c. Biceps muscle
d. Supraspinatus muscle |
|
Definition
|
|
Term
[image]
The structure represented by number 4 corresponds to the:
a. Acromion process
b. Glenoid process
c. Coracoid process
d. Lesser tuberosity of the humerus |
|
Definition
|
|
Term
The optimum visualization of the cervical roots within the intervertebral foramina is best demonstrated in a magnetic resonance image obtained in:
a. Coronal section
b. Angled oblique section
c. Parasagittal sections
d. Transverse sections |
|
Definition
|
|
Term
The principal advantage of magnetic resonance imaging compared to computed tomography
is in its ability to acquired images:
a. With a greater spatial resolution
b. In a shorter period of time
c. That is free of magnification distortion
d. In virtually any imaging plane |
|
Definition
d. In virtually any imaging plane |
|
|
Term
Prior to obtaining a magnetic resonance image in the axial plane of the lumbar spine, a localization scan is obtained in the sagittal plane to help produce image sections that are:
a. Perpendicular to the intravertebral foramina
b. Parallel to the vertebral bodies
c. Perpendicular to the intervertebral disks
d. Parallel to the spinal nerves |
|
Definition
b. Parallel to the vertebral bodies |
|
|
Term
The longitudinal extent of a lesion or pathological process can be best evaluated on
magnetic resonance images obtained in the:
1. Axial plane
2. Sagittal
3. Coronal plane
a. 1 & 2 only
b. 1 & 3 only
c. 2 & 3 only
d. 1, 2, & 3 |
|
Definition
|
|
Term
The use of magnetic resonance imaging of the hip has proved beneficial for the early detection of an evaluation of:
1. Osteonecrosis
2. Arthritis
3. Joint effusion
a. 1 only
b. 2 only
c. 3 only
d. 1, 2, & 3 |
|
Definition
|
|
Term
The most common indication of intracranial pathology on magnetic resonance images of the brain obtained in the axial plane is most often a
a. Loss of left-to-right symmetry
b. Loss of echogenicity
c. Uniform gain of signal strength
d. Loss of homogeneity of the gray matter |
|
Definition
a. Loss of left-to-right symmetry |
|
|
Term
This magnetic resonance image is best describe as a:
a. Coronal image of the elbow
b. Sagittal image of the knee*
c. Axial images of the hip
d. Off-axial image of the knee |
|
Definition
|
|
Term
[image]
This magnetic resonance image is best describe as
a:
a. Coronal image of the elbow
b. Sagittal image of the knee
c. Axial images of the hip
d. Off-axial image of the knee |
|
Definition
b. Sagittal image of the knee |
|
|
Term
[image]
This magnetic resonance image was most likely acquired using the following imaging parameters:
a. Repetition time of 200 milliseconds and an echo time of 20 milliseconds
b. Repetition time of 400 milliseconds and an echo time of 40 milliseconds
c. Repetition time of 3100 milliseconds and an echo time of 15 milliseconds
d. Repetition time of 4000 milliseconds and an echo time of 50 milliseconds |
|
Definition
c. Repetition time of 3100 milliseconds and an echo time of 15 milliseconds |
|
|
Term
[image]
On the magnetic resonance image the structure labeled number 1 corresponds to the:
a. Tendon of the quadriceps muscle
b. Anterior cruciate ligament
c. Infrapatellar fat pad
d. Tendon of the semimembranosus muscle |
|
Definition
a. Tendon of the quadriceps muscle |
|
|
Term
[image]
On the magnetic resonance image the posterior cruciate ligament is labeled:
a. Number 2
b. Number 4
c. Number 5
d. Number 9 |
|
Definition
|
|
Term
Most paramagnetic contrast agents containing gadolinium are associated with:
a. A decrease in the T1 relaxation time of the target tissue
b. An increase in the proton density of surrounding tissues
c. An increase in the T2 relaxation time of the target tissues
d. Little of or change in the factors that affect the intensity of the target tissues |
|
Definition
a. A decrease in the T1 relaxation time of the target tissue |
|
|
Term
Magnetic resonance imaging of the upper abdomen for the biliary and alimentary tract often provides little additional diagnostic information because of the:
a. A smaller field of view
b. Motion that is present in the region
c. High interference noise produced in the region
d. Chemical shift artifacts formed in the region |
|
Definition
b. Motion that is present in the region |
|
|
Term
Magnetic resonance imaging of the upper abdomen for the biliary and alimentary tract often provides little additional diagnostic information because of the:
a. A smaller field of view
b. Motion that is present in the region
c. High interference noise produced in the region
d. Chemical shift artifacts formed in the region |
|
Definition
b. Motion that is present in the region |
|
|
Term
In a T1 weighted imaging sequence, the high contrast seen on magnetic resonance images of the kidney and urinary bladder are related to:
1. Characteristic of the external tissue layers
2. Presence of urine within the organs
3. The surrounding layer of fat
a. 1 only
b. 2 only
c. 3 only
d. 1, 2 & 3 |
|
Definition
|
|
Term
In a T1 weighted imaging sequence, the high contrast seen on magnetic resonance images of the kidney and urinary bladder are related to:
1. Characteristic of the external tissue layers
2. Presence of urine within the organs
3. The surrounding layer of fat
a. 1 only
b. 2 only
c. 3 only
d. 1, 2 & 3 |
|
Definition
|
|
Term
[image]
The image plane which is best suited for the evaluation of the spinal cord and associated soft tissue structures is the:
a. True axial plane
b. Tilted axial plane
c. Coronal plane
d. Sagittal plane |
|
Definition
|
|
Term
A magnetic resonance imaging study is to be obtained of the brain. The use of thin continuous section is most often requested for patients with:
a. A sudden loss of memory
b. a sudden loss of visual acuity
c. A suspected acoustic neuroma
d. A suspected gliomas |
|
Definition
c. A suspected acoustic neuroma |
|
|
Term
The dark region appearing on magnetic resonance images made at the level of the great vessels and their entry points with the heart are most often caused by:
a. Flow void artifacts
b. Rephrasing effects
c. Moire effects
d. Parallel ring effects
|
|
Definition
|
|
Term
The structures which are well visualized on magnetic resonance images of the retroperitoneum include:
1. abdominal aorta
2. Inferior vena cava
3. Adrenal glands
a. 1 only
b. 2 only
c. 3 only
d. 1, 2 & 3 |
|
Definition
|
|
Term
During a magnetic resonance imaging of the knee with the patient supine, the anterior cruciate ligament can be placed parallel to the sagittal plane by employing a:
a. 10-15 degree internal rotation
b. 10-15 degree external rotation
c. 25-30 degree internal rotation
d. 45-50 degree external rotation |
|
Definition
b. 10-15 degree external rotation |
|
|
Term
[image]
This magnetic resonance image is best described as:
a. An axial image of the brain
b. A coronal image of the brain
c. A sagittal image of the brain
d. An oblique image of the brain |
|
Definition
c. A sagittal image of the brain |
|
|
Term
[image]
This magnetic resonance image of the brain was obtained using:
a. T1 weighted imaging sequence
b. Magnetic resonance angiographic technique
c. Spectrographic imaging sequence
d. T2 weighted imaging sequence |
|
Definition
a. T1 weighted imaging sequence |
|
|
Term
The pituitary gland is represented by number:
a. 2
b. 4
c. 7
d. 9 |
|
Definition
|
|
Term
[image]
The pituitary gland is represented by number:
a. 2
b. 4
c. 7
d. 9 |
|
Definition
|
|
Term
The structure represented by number 9 corresponds to the:
a. Corpus callosum
b. Lateral ventricle
c. Cerebellum
d. Pons
|
|
Definition
|
|
Term
The majority of the paramagnetic contrast agents used in magnetic resonance imaging are cleared from the body by a process termed:
a. Tubular reabsorption
b. Microvilli absorption
c. Glomerular filtration
d. Interstitial filtration |
|
Definition
|
|
Term
The natural contrast demonstrated on all magnetic resonance images of the pelvis is in large
part due to the:
a. Perivisceral fat of the bowel and pelvic organs
b. Tissues characteristics of intrapelvic lesions
c. High fluid content of the pelvic structures
d. Cancellous and cortical bone |
|
Definition
a. Perivisceral fat of the bowel and pelvic organs |
|
|
Term
During magnetic resonance imaging of the orbit, optimum images are obtained when the:
a. Eye continuously remains in motion
b. The eye is focused on a single direction
c. Eyes are rotated in a circular pattern
d. Eyelid are opened and closed |
|
Definition
b. Eye is focused to a single direction |
|
|
Term
During magnetic resonance imaging of the orbit, optimum images are obtained when the:
a. Eye continuously remains in motion
b. The eye is focused on a single direction
c. Eyes are rotated in a circular pattern
d. Eyelid are opened and closed |
|
Definition
b. The eye is focused to a single direction |
|
|
Term
Magnetic resonance imaging is performed using a focus or surface coils. These are most
often used to enhance the spatial resolution of the image acquired using short TR/TE spin
echo sequences for all the following EXCEPT:
a. Thorax
b. Shoulder
c. Knee
d. Pelvis |
|
Definition
|
|
Term
Improved visualized of the pre-epiglottic space and anterior commissure on a magnetic resonance image of the brain is obtained in the:
a. Coronal plane
b. Axial plane
c. sagittal plane
d. Off-axial plane |
|
Definition
|
|
Term
[image]
This magnetic resonance image of the brain was obtained in a/an:
a. Coronal plane
b. Axial plane
c. Sagittal plane
d. Off-axial plane |
|
Definition
|
|
Term
[image]
This magnetic resonance of the brain image was most likely acquired using the following imaging parameters:
a. Repetition time 600 milliseconds and an echo time 30 milliseconds
b. Repetition time 1000 milliseconds and an echo time 40 milliseconds
c. Repetition time 800 milliseconds and an echo time 15 milliseconds
d. Repetition time 2000 milliseconds and an echo time of 800 milliseconds |
|
Definition
d. Repetition time 2000 milliseconds and an echo time of 800 milliseconds |
|
|
Term
[image]
The choroid plexus which is responsible for the production of cerebrospinal fluid is labeled:
a. Number 1
b. Number 4
c. Number 6
d. Number 8 |
|
Definition
|
|
Term
[image]
The structure represented by number 3 corresponds to the:
a. Medulla oblongata
b. Thalamus
c. Corpus callosum
d. Pons |
|
Definition
|
|
Term
Magnetic resonance images of the brain are to be obtained using a series of coronal sections. Which of the following planes transects the paracentral lobule, cingulate gyrus, posterior lateral ventricle, and temporal lobe?
a. Mid-thalamic plane
b. Mid-frontal plane
c. Orbital plane
d. The plane of the pineal body |
|
Definition
|
|
Term
The normal imaging planes obtained during a magnetic resonance image of the knee to
evaluate the soft tissues damage will normally include:
1. Sagittal planes
2. Axial planes
3. Coronal planes
a. 1 & 2 only
b. 1 & 3 only
c. 2 & 3 only
d. 1,2 & 3 |
|
Definition
|
|
Term
During a magnetic resonance imaging study of the brain, particular attention to the anterolateral surface of the frontal lobe should be made in a patient experiencing:
a. An intermittent loss in their hearing
b. A loss of equilibrium
c. A disturbance in their speech
d. A disturbance in their vision |
|
Definition
a. An intermittent loss in their hearing |
|
|
Term
The low-intensity signal from the cerebral spinal fluids within the ventricles of the brain is normally the results of:
a. An inversion recovery artifacts
b. A flow void artifacts
c. A paramagnetic artifacts
d. A recoil echo artifacts |
|
Definition
|
|
Term
During a magnetic resonance imaging procedure the use of a whole volume quadrative(saddle) coil has shown to be beneficial for the improved visualization of the structure in the:
a. Upper abdominal cavity
b. Shoulder and the knee
c. The lower abdominal cavity and pelvis
d. The cervical region of the spine |
|
Definition
d. The cervical region of the spine |
|
|
Term
This magnetic resonance of the lumbar spine was acquired in the:
a. Axial plane
b. Sagittal plane
c. Coronal plane
d. Off-axial plane |
|
Definition
|
|
Term
[image]
This magnetic resonance image was obtained using a 1-tesla unit. This image was most likely acquired using the following parameters:
a. A repetition time of 2000 milliseconds and an echo time of 20 milliseconds
b. A repetition time of 2000 milliseconds and an echo time of 40 milliseconds
c. A repetition time of 1500 milliseconds and an echo time of 80 milliseconds
d. A repetition time of 500 milliseconds and an echo time of 20 milliseconds |
|
Definition
d. A repetition time of 500 milliseconds and an echo time of 20 milliseconds |
|
|
Term
[image]
The structure represented by number 3 corresponds to the:
a. Vertebral foramen
b. Spinalis muscle
c. Conus medullaris
d. Vertebral body |
|
Definition
|
|
Term
[image]
The intervertebral disk corresponds to:
a. Number 1
b. Number 2
c. Number 3
d. Number 5 |
|
Definition
|
|
Term
During a magnetic resonance imaging study of the thoracic or lumbar spine, a field of view better approximating the region of interest can be accomplished by the placement of the planar
surface coil:
a. With its long axis parallel to the spine
b. With its long axis perpendicular to the spine
c. With its long axis oblique to the spine
d. With its long axis tangent to the spine |
|
Definition
a. With its long axis parallel to the spine |
|
|
Term
Prior to a magnetic resonance imaging study of the female pelvis, the administration of
glucagon can be useful to:
a. Reduce the superimposition of overlaying structures
b. Reduce the production of flow void artifacts
c. Eliminate residual fluid from the urinary bladder
d. Reduce peristalsis in the bowel |
|
Definition
d. Reduce peristalsis in the bowel |
|
|
Term
During MR imaging of the head what structure is likely to demonstrate using a series
of coronal slices that extend from the anterior to posterior clinoid process:
a. The pituitary gland
b. The Circle of Willis
c. The internal auditory canals
d. Ocular mass lesion
|
|
Definition
|
|
Term
MR imaging of the tissues of the neck should be obtained for all of the following EXCEPT:
a. The staging of laryngeal carcinomas
b. The assessment salivary gland neoplasms
c. Assessment od paraphryngeal masses
d. The evalation of brain stem tumors |
|
Definition
d. The evalation of brain stem tumors |
|
|
Term
During MR imaging of the cervical spine, the administration of a paramagnetic contrast agent
is likely to enhance all of the following EXCEPT:
a. Postsurgical scar tissue
b. Recurrent disk disease
c. Metastatic lesion
d. Multiple sclerosis plaques |
|
Definition
b. Recurrent disk disease |
|
|
Term
The lower signal intensity associated with long echo time (TE) is likely to result in all the following EXCEPT:
a. A higher spatial resolution
b. A higher amount of image graininess
c. A lower signal intensity
d. A lower signal-to-noise ratio |
|
Definition
d. A lower signal-to-noise ratio |
|
|
Term
The introduction of paramagnetic substances such as gadolinium on MR images of the brain are often used to enhance:
a. Active hemorrhages
b. Meningiomas
c. Subdural hematomas
d. Hydrocephalus |
|
Definition
|
|
Term
In addition to its effects on the contrast of the image, the repetition time (TR) during MR imaging, is most closely related to a/an:
a. Increase in the thickness of the slice
b. Modification to the orientation of the images plane
c. Reduction in the acquisition time
d. Increase in the field of view |
|
Definition
c. Reduction in the acquisition time |
|
|
Term
The minimum echo time(TE) used for most modern MR imaging studies is in the range of:
a. 3-5 milliseconds
b. 5-15 milliseconds
c. 30-40 milliseconds
d. 50-70 milliseconds |
|
Definition
|
|
Term
A 90° radiofrequency (RF) pulse derives its names from the:
a. The direction of the emission pulse relative to the main magnet
b. Number of degrees of the field
c. Frequency of the pulse that is required to tilt a given hydrogen atom
d. Degree of tilting the pulse causes in the precessing protons |
|
Definition
b. Number of degrees of the field |
|
|
Term
Each slice of the MR image consists of a series of volume elements called:
a. Voxels
b. Pixels
c. Matrices
d. Vectors |
|
Definition
|
|
Term
Which of the following tissues is most often associated with a high signal intensity and a shortened T1 relaxation time during both T1 and T2 MR images?
a. Cortical bone
b. The spinal cord
c. Fat or cholesterol
d. Cerebral fluid |
|
Definition
|
|
Term
A kinematic(cine)MR study in which multiple MR images are obtained and displayed in rapid
succession is most often used to:
a. Provided a 3-Dimensional image of the abdomen
b. Resemble the phases of respiration
c. Demonstrate the speed of blood flow
d. Resemble the motion of a joint |
|
Definition
d. Resemble the motion of a joint |
|
|
Term
The intensity value of a pixel in an MR image display is dependent upon the:
a. Signal intensity from the entire voxel
b. Signal intensity from filtered voxels
c. Signal intensity from multiple voxels
d. The thickness of the tissue volume |
|
Definition
c. Signal intensity from multiple voxels |
|
|
Term
Which of the following is the most likely value for a long echo time?
a. 10 milliseconds
b. 500 milliseconds
c. 100 milliseconds
d. 30 milliseconds |
|
Definition
|
|
Term
Which of the following is an MR imaging technique that can be used to determine the relative quantity of a chemical substance in a given volume of tissue?
a. Magnetic resonance spectrography
b. Magnetic resonance subtraction
c. Phase-contrast magnetic resonance
d. Temporal flow magnetic resonance |
|
Definition
d. Temporal flow magnetic resonance |
|
|
Term
Which of the following represents the correct sequence for an end spin echo pulse sequence?
1. Signal measurement
2. 180-degree pulse
3. 90-degree pulse
a. 123
b. 213
c. 321
d. 132 |
|
Definition
|
|
Term
During an MR imaging procedure, a high contrast low grain image is normally associated with all following EXCEPT:
a. A lower signal-to-noise ratio and lower field strength
b. Lower field strength and longer bandwidth
c. A lower radiofrequency pulses
d. Higher field strength and shorter bandwidth |
|
Definition
d. Higher field strength and shorter bandwidth |
|
|
Term
The slowing in the velocity of the arterial blood flow during the production of flow sensition images during MR angiography is most often accomplished by:
a. Gating of the pulses to the diastolic phase
b. Reduction of the cardiac flow rate
c. Slowing the contraction rate of the heart
d. The administration of beta-blocking drugs
|
|
Definition
a. Gating of the pulses to the diastolic phase |
|
|
Term
The principal reason that the T2 signal must be detected a short time after the 90-degree pulse is turned off is:
1. Noise levels tend to increase with time
2. Signal intensity decreases with time
3. Transverse magnetization increases within 80 milliseconds of the pulse
a. 1 only
b. 2 only
c. 3 only
d. 1,2&3 |
|
Definition
|
|
Term
When the protons densities of tissue are nearly equal, the highest intensity signals are normally
detected from the substances with:
a. Short T1 times
b. Long T1 times
c. Greatest tissue movement
d. Shortest T2 times |
|
Definition
|
|
Term
All of the following images or equipment controls must be available to the operator of an MR imaging
unit EXCEPT:
a. The slice thickness
b. The imaging plane orientation
c. The size of the image matrix
d. The main magnet power control |
|
Definition
d. The main magnet power control |
|
|
Term
The changes in the signal intensity due to different locations of the echos from moving tissue is termed the:
a. Gating time effect
b. Doppler effect
c. Time-of-flight effect
d. Spin echo effect |
|
Definition
|
|
Term
The signal intensity of the T2 curve is most closely related to the energy exchange between protons called transverse relaxation and is most dependent upon the:
a. Electron density of the matter
b. Chemical environment of the tissue
c. Nuclear stability of the tissue
d. Physical density of the tissue |
|
Definition
b. Chemical environment of the tissue |
|
|
Term
Which of the following flip angles is associated with the release of the least amount of transverse magnetization?
a. 180 degrees
b. 90 degrees
c. 60 degrees
d. 30 degrees |
|
Definition
|
|
Term
During a spin echo pulse sequence, a 180° pulse that is applied a short time after the 90° pulse will serve to:
a. Compensate for non-uniformities in the protons release of transverse magnetism
b. Reverse the spin of the nuclear protons and release longitudinal magnetism
c. Initiate protons to come into phase and release longitudinal magnetism
d. Enable protons to regain their phase and recover transverse magnetization
|
|
Definition
d. Enable protons to regain their phase and recover transverse magnetization
|
|
|
Term
After the raw data has been manipulated to form the computer-constructed images this information is most often stored on:
a. Magnetic tapes or optical disks
b. Integrated circuits chips
c. Liquid crystal storage systems
d. Luminescent crystals |
|
Definition
a. Magnetic tapes or optical disks |
|
|
Term
The principal advantage of an MR angiography compared to radiographic angiography techniques
is that the:
a. MR imaging techniques are less expensive
b. MR imaging techniques have a higher spatial resolution
c. MR imaging techniques provide for an improved vessel selectivity
d. MR imaging techniques are non-invasive
|
|
Definition
d. MR imaging techniques are non-invasive |
|
|
Term
A decrease in the spatial resolution of the MR image due to increase in the field of view(FOV) is most often
associated with:
a. An increase in the size of the pixels
b. A decrease in the signal-to-noise ratio
c. An increase in the size of the matrix
d. A decrease in the size of the monitor |
|
Definition
a. An increase in the size of the pixels |
|
|
Term
The process by which a staggered pulse sequence is employed to shorten the imaging time of a given region is called:
a. Flip angle imaging
b. Multislice imaging
c. Gradient echo imaging
d. Gated imaging |
|
Definition
|
|
Term
A 180°-190° inversion recovery pulse pair is most often employed for the production of a/an:
a. Multi-slice decay signal
b. Reynold's decay signal
c. Free induction decay signal
d. Inter-pulse decay signal |
|
Definition
c. Free induction decay signal |
|
|
Term
The bandwidth of the radiofrequency (RF) pulse for a given field strength of an MR image is most closely related to the:
a. Length of the image acquisition
b. Strength of the field
c. Repetition time of the sequence
d. Thickness of the slice |
|
Definition
d. Thickness of the slice |
|
|
Term
In a spin echo technique, the time between the initial 90° pulse and terminal 180° pulse is called:
a. Echo time (TE)
b. Inversion time (TI)
c. Repetition time (TR)
d. Spin time (TS) |
|
Definition
|
|
Term
During MR angiography, the suppression of static (non-moving) tissues is most often accomplished by the use of:
a. Subtraction techniques
b. Registration averaging
c. Gradient reversal techniques
d. Flow compensation |
|
Definition
a. Subtraction techniques |
|
|
Term
[image]
Which of the following images of the spine was acquired using a longer echo time (TE)
a. Image 1
b. Image 2 |
|
Definition
|
|
Term
[image]
The image represented by number 2 can best be described as a:
a. T1 weighted image
b. Proton density
c. T2 weighted image
d. Spectrographic image |
|
Definition
|
|
Term
Time-of-flight effects are primarily dependent upon:
a. Metabolism of glucose by liver
b. Absorption of fluids by the villi
c. Peristalsis of the GI tract organs
d. Movement of the blood within the vessel |
|
Definition
d. Movement of the blood within the vessel |
|
|
Term
The contrast produced during a heavy T! weighted sequence using a short echo time (TE) and short
repetition time (TR) is inversely related to the:
a. TE/T1 ratio
b. T2/TR ratio
c. TR/T1 ratio
d. T2/TE ratio |
|
Definition
|
|
Term
How many 180° RF pulses are used during each TR period in a signal-slice GRASE sequence which collects 4 spin echoes each surrounded by 8 gradient echoes:
a. 4
b. 8
c. 16
d. 32 |
|
Definition
|
|
Term
Using a spin-echo pulse sequence which generates two echoes, imagine a hypothetical tissue in which the protons never experience spin-spin interactions. How does the signal from the first echo compare to the signal from the second echo?
a. Its frequency is higher
b. Its frequency is lower
c. It is the same size
d. It is smaller in size |
|
Definition
|
|
Term
[image]
In the image, which arrow is pointing at the ulna:
a. a
b. b
c. c
d. d |
|
Definition
|
|
Term
Which of these flow effects cause an increase in signal intensity?
1. Entry phenomenon
2. Even echo rephasing
3. Diastolic pseudogating
a. 1 only
b. 1 and 2 only
c. 1 and 3 only
d. 1, 2, and 3 |
|
Definition
d
All of these flow effects exhibit increased signal intensity |
|
|
Term
What is another name for thermal relaxation?
a. Spin-lattice relaxation
b. Spin-spin relaxation
c. Spin density relaxation
d. None of the above
|
|
Definition
a
T1 relaxation hastens with Brownian motion therefore thermal relaxation is sometimes used to describe spin-lattice relaxation |
|
|
Term
What is the scan time of a dual contrast spin echo pulse sequence with a TR of 2500 msec, 150 phase steps in the matrix and 1 signal averages:
a. 3.65 minutes
b. 4,27 minutes
c. 6.25 minutes
d. Not enough information is given |
|
Definition
|
|
Term
The advantages of rapid scanning techniques include all the following except:
a. Improved patient comfort
b. The possibility of producing subtracted images
c. Reduction of patient motion
d. The possibility of dynamic imaging |
|
Definition
b. The possibility of producing subtracted images |
|
|
Term
Which image weighting is likely to occur from the use of a 2000 millisecond (long) repetition time (TR) and a minimal (20 millisecond) echo time (TE)?
a. Spin-lattice relaxation (T1)
b. Spin-spin relaxation (T2)
c. Proton density
d. Reverse spin density |
|
Definition
|
|
Term
Which of the following statements is true of three-dimensional Fourier transform (3-DFT) image acquisition compared to a two-dimensional Fourier transform?
a. Data is acquired from the entire imaging volume
b. The pulse repetition time is lengthened
c. A narrow-banded excitation pulse is employed
d. Image resolution is substantially improved |
|
Definition
a. Data is acquired from the entire imaging volume |
|
|
Term
which of the following is the example of an inversion recovery (IR) spin echo sequence?
a. 90°pulse - 180°pulse - 180°pulse
b. 90°pulse - 90°pulse - 90°pulse
c. 180°pulse - 90°pulse - 180°pulse
d. 180°pulse - 90°pulse - 90°pulse |
|
Definition
c. 180°pulse - 90°pulse - 180°pulse |
|
|
Term
Calculate the total scan time for an image obtained using a repetition time (TR) of 1000 milliseconds, 128 lines and 3 excitations.
a. 6.4 minutes
b. 4.2 minutes
c. 2.1 minutes
d. 1 minutes |
|
Definition
|
|
Term
The first step in the production of MR angiographic image is a projection image which can be obtained from a:
a. Single thin slices 2-DFT image
b. Single thick 3-DFT
c. Background suppressed (subtraction) image
d. Flow-sensitive (gated) image |
|
Definition
|
|
Term
The human eye is most likely to detect a substantial loss of contrast on MR images obtained at:
a. High signal-to-noise levels and higher field strengths
b. High signal-to-noise levels and shorter radiofrequency bandwidth
c. Shorter radiofrequency bandwidth and lower field strengths
d. Lower signal-to-noise levels and higher radiofrequency bandwidth |
|
Definition
d. Lower signal-to-noise levels and higher radiofrequency bandwidth |
|
|
Term
Flow sensitization in MR angiography can be improved by all the following EXCEPT:
a. Gating techniques
b. Pulse averaging
c. Phase shifting
d. Time of flight techniques |
|
Definition
|
|
Term
A heavily weighted spin-spin relation or T2 image can be obtained by which of the following repetition and echo time combinations?
a. 2000 millisecond TR and a 20 millisecond TE
b. 300 millisecond TR and a 20 millisecond TE
c. 2000 millisecond TR and a 70 millisecond TE
d. 800 millisecond TR an a 40 millisecond TE |
|
Definition
c. 2000 millisecond TR and a 70 millisecond TE |
|
|
Term
The signal-to-noise ratio in a given pixel is proportional to all the following EXCEPT:
a. Volume of the voxels used in the image display
b. Square root of the number of excitations
c. Square root of the number of radiofrequency pulse
d. Square root of the number of phase encoding projections |
|
Definition
d. Square root of the number of phase encoding projections |
|
|
Term
A shortening of the echo sampling time and an increase in the T1 weighting is associated with
a. An increased image contrast
b. An increased image noise
c. An increased spatial resolution
d. An increased field of view |
|
Definition
b. An increased image noise |
|
|
Term
A large volume excitation pulse in conjunction with a shortened repetition pulse is most commonly
used in the acquisition of:
a. Any image plane
b. Large field of view images
c. Multiple slice reconstruction
d. Multi-planer images |
|
Definition
|
|
Term
The excitation of tissues outside the slice of tissue can result in crosstalk and the perception that the thickness of the slice has been increased. This effect is most easily avoided by the use of:
a. Magnetic collimators
b. Separations of gaps between the slices
c. Surfaces coils
d. Reductions in the thickness of the slices |
|
Definition
b. Separations of gaps between the slices |
|
|
Term
The T2 is defined as the time it takes to decrease the transverse magnetism to:
a. 17% of its original value
b. 37% of its original value
c. 63% of its original value
d. 79% of its original value |
|
Definition
b. 37% of its original value |
|
|
Term
During an inversion recovery sequence the signal intensity is most closely related to the:
a. T1 relaxation
b. T2 relaxation
c. Proton density
d. Spin density |
|
Definition
|
|
Term
In order to provide reduced motion in cardiac studies, the images are gated to each heartbeat.
The trigger is normally the:
a. End of the respiratory cycle
b. Systolic pulse of the brachial artery
c. R wave of the EKG
d. Middle of the QRS complex |
|
Definition
|
|
Term
What is the maximum number of slices permitted for a spin echo sequence with a repetition time of 500 milliseconds and an echo time of 50 milliseconds? Assuming a 10 millisecond constant.
a. 6
b. 12
c. 16
d. 20 |
|
Definition
|
|
Term
Pertaining to the information in question 126, what slice thickness should be selected if the entire organ of interest measures 23 centimeters?
a. 1 centimeter slices
b. 1.5 centimeter slices
c. 2 centimeter slices
d. 4 centimeter slices |
|
Definition
|
|
Term
The main advantage of an MR imaging compared to either conventional radiography or computed
tomography is a/an:
a. Higher spatial resolution
b. Improve low contrast resolution
c. Reduced image magnification
d. Reduced patient motion |
|
Definition
b. Improve low contrast resolution |
|
|
Term
All of the following image parameters will have an effect on the spatial resolution of an MR image EXCEPT:
a. The size of the image matrix
b. The field of view
c. The gradient field strength
d. The time of flight |
|
Definition
|
|
Term
During MR imaging a repetition of the imaging sequence and/or the averaging of the signals are most often used to:
a. Increase the signal-to-noise ratio (SNR)
b. Improve spatial resolution
c. Decrease the signal intensity
d. Suppress background signal |
|
Definition
a. Increase the signal-to-noise ratio (SNR) |
|
|
Term
Which of the following statements is true concerning the effects of echo time (TE) between tissues possessing T2 and short T2 values?
a. T2 contrast is maximized at a short echo time (TE)
b. The signal-to-noise ratio is highest with a long time (TE)
c. T2 contrast is maximized at a long echo time (TE)
d. T2 has little effect on the appearance of different tissues |
|
Definition
c. T2 contrast is maximized at a long echo time (TE) |
|
|
Term
A decrease in the bandwidth of the radiofrequency (RF) pulse during the acquisition of an MR image is likely to result in:
a. An increase in the signal-to-noise ratio
b. An increase in the field of view
c. An increase in the images acquisition time
d. A reduction in the number of image artifacts |
|
Definition
a. An increase in the signal-to-noise ratio |
|
|
Term
During an MR imaging procedure, when the strength of the gradient field is low, an artifact between fat and water may occur. This is referred to as:
a. An intensity coefficient artifact
b. A chemical shift artifact
c. A hydrodynamic artifact
d. A signal averaging artifact
|
|
Definition
b. A chemical shift artifact |
|
|
Term
All of the following are classified as interpulse delay times of a spin echo sequence EXCEPT:
a. Subversion time (TS)
b. Repetition time (TR)
c. Inversion time (TI)
d. Echo time (TE) |
|
Definition
|
|
Term
What is the minimum repetition time(TR) that would be required for a gated cardiac magnetic imaging procedure for a patient that has a heart rate of 120 beats per minute?
a. 50 milliseconds
b. 250 milliseconds
c. 500 milliseconds
d. 1000 milliseconds
|
|
Definition
|
|
Term
The recommended dosage of gadoteridol or gadopentetate dimeglumine injections is:
a. .1 mmol/kg (.2 ml/kg)
b. .5 mmol/kg (1 ml/kg)
c. 1 mmol/kg (2 ml/kg)
d. 3 mmol/kg (6 ml/kg) |
|
Definition
|
|
Term
MR imaging for the central nervous system is normally superior to computed tomographic images for the diagnosis of:
a. Neoplastic lesions
b. Arteriovenous malformations
c. Inflammatory conditions
d. Acute hemorrhage |
|
Definition
|
|
Term
Calculate the pixel size for a 10-centimeter field of view and a 256x256 image matrix
a. 6.5 millimeters
b. 13 millimeters
c. 25.6 millimeters
d. 39 millimeters |
|
Definition
|
|
Term
All magnetic fields are to some extent non-uniform. Which of the following devices can be used to increase this uniformity?
a. Shim coils
b. Whole volume coils
c. Surface coils
d. Quadrature coils |
|
Definition
|
|
Term
In the vast majority of modern MR imaging units, the main magnetic field is provided by:
a. Resistive magnets
b. Permanent magnets
c. Superconducting magnets
d. Electrothermal magnets |
|
Definition
c. Superconducting magnets |
|
|
Term
The spinning of the proton in the hydrogen nucleus is the most closely related to the:
a. Generation of radiofrequencies (RF) pulses by the atom
b. Generation of time-varying magnetism in the atom
c. Generation of magnetism or magnetic moments by the atom
d. Generation of electromagnetic disturbances in the atom
|
|
Definition
c. Generation of magnetism or magnetic moments by the atom |
|
|
Term
During an MR imaging procedure, in order for the tissues to be magnetized in the longitudinal direction, the main magnetic field should be aligned with the:
a. Long axis of the patient (Z-axis)
b. Transverse axis of the patient (X-axis)
c. Axial axis of the patient (Y-axis)
d. Oblique place of the patient (O-axis) |
|
Definition
a. Long axis of the patient (Z-axis) |
|
|
Term
The basic principles of MR were in large part due to research performed by:
a. Faraday and Ohms
b. Edison an Oersted
c. Crick and Watson
d. Bloch and Purcell
|
|
Definition
|
|
Term
The magnetic component of the radiofrequency coils, which provide the means for inducing the magnetic resonance in the tissue, must be capable of producing:
a. A single range of frequencies
b. A narrow range of frequencies
c. A broad range of frequencies
d. Only Lamor frequencies |
|
Definition
c. A broad range of frequencies |
|
|
Term
In a uniform magnetic field the shortening of the echo time (TE) is most commonly associated with production of a:
a. Gyro-magnetic artifact
b. Field-induction decay artifact
c. Precession delay artifact
d. Cross-talk artifact |
|
Definition
b. Field-induction decay artifact |
|
|
Term
Which of the following devices is associated with a reduction in the field of view and an improved signal-to-noise ratio during MR imaging?
a. The application of surface coils
b. The use of ferromagnetic shielding coils
c. The application of Lamor frequency enhancers
d. The application of internal magnetic implants |
|
Definition
a. The application of surface coils |
|
|
Term
The precession frequency(wobble) of a given atom is related to all the following EXCEPT:
a. The temperature of the atom
b. The number of nuclear protons
c. The strength of the magnetic field
d. The chemical bonding of the atoms |
|
Definition
a. The temperature of the atom |
|
|
Term
The type of magnetism produced when the net magnetism of the field is modified by a radiofrequency (RF) pulse is termed:
a. Gradient magnetism
b. Transverse magnetism
c. Longitudinal magnetism
d. Alternating magnetism |
|
Definition
|
|
Term
In a spin echo imaging sequence, the slice selection gradient in a conventional two-dimensional Fourier transform imaging sequence is applied between the:
a. 180°radiofrequency pulses and the 90° radiofrequency pulses
b. 60° radiofrequency pulses and the 90° radiofrequency pulses
c. 90° radiofrequency pulses and the 180°radiofrequency pulses
d. 270°radiofrequency pulses and the 90°radiofrequency pulses |
|
Definition
c. 90° radiofrequency pulses and the 180°radiofrequency pulses |
|
|
Term
Quadrature coils which are specifically designed to produce a circular polarization of radio waves are commonly employed during an MR imaging procedure to do all of the following EXCEPT:
a. Decrease the heating in the tissues
b. Focus the radio waves into a thin plane
c. Improve the radiofrequency homogeneity
d. Improve the signal-to-noise ratio |
|
Definition
b. Focus the radio waves into a thin plane |
|
|
Term
During an MR imaging procedure an axial gradient will tend to:
a. Increase the resonant frequency of the photons nearest the head of the patient
b. Increase the resonant frequency of the photons at the lateral regions of the field
c. Equalize the resonant frequency of the photons at the head and feet of the patient
d. Equalize the resonant frequency of the photons at the front and back of the patient
|
|
Definition
a. Increase the resonant frequency of the photons nearest the head of the patient |
|
|
Term
The movement of a nucleus about its axis in a fashion similar to a spinning top in a strong external magnetic field is termed:
a. Relaxation
b. Parallelism
c. Precession
d. Synchronism |
|
Definition
|
|
Term
The Larmor equation w=g Bo is used in the determination of the:
a. Direction of the phase pulse
b. T1 relaxation time
c. Transverse relaxation time
d. Precession frequency |
|
Definition
|
|
Term
During an MR imaging the net magnetic vector is derived from:
a. The K shell electron location in the hydrogen atoms
b. Approximately half of the protons located in the hydrogen atoms
c. A small percentage of the protons located in the hydrogen atoms
d. All the protons located in the hydrogen atoms |
|
Definition
c. A small percentage of the protons located in the hydrogen atoms |
|
|
Term
An image, that has an area of geometric distortion due to non-uniformities of the magnetic field, is often caused by:
a. Misaligned surface coils
b. Ferromagnetic objects in the field
c. Non-magnetic objects in the field
d. Magnetic anisotropy |
|
Definition
b. Ferromagnetic objects in the field |
|
|
Term
During a magnetic resonance imaging procedure the high magnetic field strength is found within the:
a. Main magnetic coils
b. Shim coils
c. Gradient coils
d. Surface coils |
|
Definition
|
|
Term
Which of the following pathologic conditions is most likely to produce partial volume averaging artifacts on
magnetic resonance image obtained of a coronal image of the spine?
a. Calcified intervertebral disks
b. Rotary or lateral scoliosis
c. Kyphosis or lordosis
d. Spondylolisthesis |
|
Definition
|
|
Term
The precession frequency for a hydrogen proton (nucleus) in a two-tesla field is about:
a. 84 MHz
b. 63 MHz
c. 42 MHz
d. 21 MHz |
|
Definition
|
|
Term
During a magnetic resonance imaging procedure, the application of radio waves at the Larmor frequency will result in a processing proton that has an increased amount of:
a. Coherent magnetic and decreased amount of random magnetism
b. Axial magnetism and decreased amount of transverse magnetism
c. Longitudinal magnetism and decreased amount of axial magnetism
d. Transverse magnetism and decreased amount of longitudinal magnetism |
|
Definition
c. Longitudinal magnetism and decreased amount of axial magnetism |
|
|
Term
During MR imaging the average time it takes for the surrounding molecules to reorient themselves to the
horizontal magnetic field during spin-lattice relaxation (T1) is strongly dependent upon the:
a. Atomic number of the tissue
b. Size of the molecules in the tissues
c. Electron density of the tissue
d. Binding energy of the atoms in the tissues |
|
Definition
b. Size of the molecules in the tissues |
|
|
Term
Which of the following compounds is normally associated with the shortest T2?
a. Pure liquids
b. Impure solids
c. Pure solids
d. Impure liquids |
|
Definition
|
|
Term
During an MR imaging study, though the receive only coils do provide for improved uniformity, they will require:
a. Multiple radiofrequencies during transmission to record all signals
b. Radiofrequency filtration circuits to produce uniform signal pulses
c. Decoupling during the transmission of a radiofrequency pulse to avoid damages to coil
d. A reduction in the acquisition rate |
|
Definition
c. Decoupling during the transmission of a radiofrequency pulse to avoid damages to coil |
|
|
Term
The main disadvantage of a superconducting type magnetic resonance unit is its need for:
a. Periodic cryogen replenishment
b. Daily realignment of the phase generator
c. A high-quality external power supply
d. Weekly transformer realignments |
|
Definition
a. Periodic cryogen replenishment |
|
|
Term
During an MR imaging procedure, electronic reorientation of the radio frequency (RF) pulses is most closely related to:
a. The thickness of the tissue section
b. The plane of the tissue that is imaged
c. The direction of the static field
d. The time of the image acquisition |
|
Definition
b. The plane of the tissue that is imaged |
|
|
Term
Most patient prior to their placement is a strong magnetic field have a net magnetism of zero because:
a. The body contains few hydrogen nuclei
b. The movements of the tissues cancel the net magnetism
c. Hydrogen nuclei are randomly oriented within the tissues
d. Nuclear spins are apparent only in a strong magnetic field |
|
Definition
c. Hydrogen nuclei are randomly oriented within the tissues |
|
|
Term
In MR imaging the terms anti-parallel (spin-up) and anti-parallel (spin down) refer to the:
a. Alignment of the hydrogen nuclei with the external magnetic field
b. Process by which chemical changes effect magnetic changes
c. Random orientation of the tissues within a magnetic field
d. Resonant frequency of the hydrogen nucleus |
|
Definition
a. Alignment of the hydrogen nuclei with the external magnetic field |
|
|
Term
Which of the following will increase the signal of the spin echo during an MR imaging procedure?
a. The ionization of the hydrogen atom
b. A higher hydrogen or proton density
c. A higher atomic number of the atom
d. The chemical bonding of a compound
|
|
Definition
a. The ionization of the hydrogen atom |
|
|
Term
During an MR imaging procedure water and other pure liquids are most often associated with a:
a. Short T1 and a short T2
b. Short T1 and a long T2
c. Long T1 and a short T2
d. Long T1 and a Long T2
|
|
Definition
|
|
Term
The loss of the established transverse magnetism when a radiofrequency is turned off is normally measured as an increase in the:
a. Precession frequency
b. External field gradient
c. Longitudinal magnetism
d. Synchronization energy |
|
Definition
d. Synchronization energy |
|
|
Term
In a time versus longitudinal magnification plot, the magnetization will increase to a maximum value after the switching off of the radiofrequency pulse. This curve is referred to as the:
a. Spin echo curve
b. Spin-lattice relaxation curve
c. Spin-spin relaxation
d. Spin proton density curve |
|
Definition
b. Spin-lattice relaxation curve |
|
|
Term
The signal in an MR imaging system is derived from the moving magnetic fields of precessing protons in phase with each other and the:
a. Electric current generated in the antenna
b. Magnetic disruption of the gradient field
c. Electromagnetic pulse generated in the antenna
d. Secondary magnetism generated in the main magnetic field |
|
Definition
a. Electric current generated in the antenna |
|
|
Term
The frequency of the MR signal will normally be:
a. Higher than the precession frequency of the hydrogen atom
b. Lower than the precession frequency of the hydrogen atom
c. The same as the precession frequency of the hydrogen atom
d. The same as the precession frequency of the oxygen atom
|
|
Definition
c. The same as the precession frequency of the hydrogen atom |
|
|
Term
The ideal surface (local) coil used for an MR imaging procedure should have all the following properties EXCEPT:
a. The surface coil should be about the same size as the structure of interest
b. The surface coil should generate radio waves which are perpendicular to the static field
c. The surface coil should be placed as close to the structure of interest as possible
d. The surface coil should extend beyond the bore of the magnet |
|
Definition
d. The surface coil should extend beyond the bore of the magnet |
|
|
Term
The gradient coils which surround the inside bore the main magnet provide magnetic fields which acts:
a. To align the main static magnetic field to the radiofrequency pulses
b. To obtain positional information required for the acquisition of the image
c. As the receiving antenna for the radiofrequency (RF) pulses
d. To redirect the radiofrequency (RF) pulses to the desired anatomic region
|
|
Definition
c. As the receiving antenna for the radiofrequency (RF) pulses |
|
|
Term
The time it takes following the radiofrequency pulse to reduce transverse magnetization in a tissue is known as the:
a. T1 echo-spin redistribution curve
b. T1 echo-echo relaxation curve
c. T2 or spin-echo redistribution curve
d. T2 or spin-spin relaxation curve |
|
Definition
d. T2 or spin-spin relaxation curve |
|
|
Term
The principal effect of a radio frequency pulse on the nuclear photons of the hydrogen atom is the:
a. Cancellation of the net magnetic moments
b. Cessation of the precession frequency
c. Synchronization or phasing of their magnetic vectors
d. Induction of a new electrical field |
|
Definition
c. Synchronization or phasing of their magnetic vectors |
|
|
Term
Spin-lattice relaxation (T1) or the time takes to reorientate to the longitudinal magnetic field, is shortest for:
a. Proteins
b. Deoxyribonucleic acid
c. Albumin
d. Lipids |
|
Definition
|
|
Term
The frequency of the waves emitted by a tissue following its exposure to a static magnetic field and a radiofrequency pulse in an MR imaging procedure is termed:
a. Resonant magnetic decay
b. Free induction decay
c. Frequency echo decay
d. Lattice gradient decay |
|
Definition
|
|
Term
Which of the following tissues is most likely to be associated with a short T1?
a. Cardiac muscle
b. Cortical bone
c. Yellow bone marrow
d. Whole blood |
|
Definition
|
|
Term
Aliasing or the superimposition of reconstruction tissues over the original image is most often seen when:
a. A surface coil is misaligned to the static magnetic field
b. The bandwidth of radiofrequency pulses exceeds recommended limits
c. The concentration of a paramagnetic agent in the target organ is excessively high
d. The structure of interest extends outside of the imaging volume |
|
Definition
d. The structure of interest extends outside of the imaging volume |
|
|
Term
Which of the following would be a valid reason for using STEAM?
a. Localizing cancers and metastasis in the brain
b. Measuring choline concentration in a small volume
c. Monitoring the effects of ischemia and stroke
d. Tracing out white matter tracts along all three dimensions
|
|
Definition
b. Measuring choline concentration in a small volume |
|
|
Term
How does gadolinium increase contrast in tissues?
a. It decreases the T1 times of tissues
b. It enhances the effects of magnetic inhomogeneities
c. It increases the T2 times of tissues
d. It reduces the proton density of the tissues
|
|
Definition
a. It decreases the T1 times of tissues |
|
|
Term
What describes the relationship between the strength of the magnetic field, the gyromagnetic ratio, and the precessional frequency?
a. Faraday’s Law
b. Ohm’s Law
c. Quantum Theory
d. The Larmor equation
|
|
Definition
|
|
Term
If the flow velocity is much lower than the VENC set for a PC-MRA, what will happen to the image?
a. The image will demonstrate very slow flows to be stationary
b. The image will have a decreased field-of-view
c. The center of the vessel and the sides of the vessel will have signals of opposite intensities
d. The image will be normal
|
|
Definition
a. The image will demonstrate very slow flows to be stationary |
|
|
Term
If the receive bandwidth is increased by a factor of 4, what will happen to the SNR?
a. It decreases by a factor of 4
b. It decreases by a factor of 2
c. It increases by a factor of 2
d. It increases by a factor of 4
|
|
Definition
b. It decreases by a factor of 2 |
|
|
Term
Which of the following does NOT involve any post-processing techniques?
a. ADC map
b. Dixon technique
c. FIESTA
d. Maximum intensity projection
|
|
Definition
|
|
Term
K-space is considered to have conjugate symmetry. This means that a signal which is collected in the top left corner of k-space has a corresponding signal in the __________ of k-space:
a. Bottom left corner
b. Bottom right corner
c. Center
d. Top right corner
|
|
Definition
|
|
Term
What is another name for inversion time?
a. Flip time
b. Relaxation time
c. Saturation time
d. Tau |
|
Definition
|
|
Term
Which of the following uses a variable flip angle, a spoiler, and a rewinder to create an echo?
a. Single spin echo pulse sequence
b. Dual spin echo pulse sequence
c. Fast spin echo pulse sequence
d. Gradient echo pulse sequence |
|
Definition
d. Gradient echo pulse sequence |
|
|
Term
If two types of tissues are represented within one voxel, their signals are averaged together. What term is used to describe this phenomenon?
a. Chemical pre-saturation
b. Fractional averaging
c. Partial voluming
d. Volume imaging |
|
Definition
|
|
Term
If a voxel has dimensions of 2x2x2 mm, it is considered to be ________:
a. Rectangular
b. Square
c. Isotropic
d. Anisotropic |
|
Definition
|
|
Term
Which term describes the maximum strength of a gradient and affects quality of resolution?
a. Rise time
b. Gradient amplitude
c. Bandwidth
d. Duty cycle
|
|
Definition
|
|
Term
Which of the following improves the signal-to-noise ratio (SNR) in an RF receiver coil?
a. Using more coils
b. Using larger coils
c. Increasing the distance between the patient and the coil
d. Using less conductive loops of wire for the coil
|
|
Definition
|
|
Term
If only 75% of K space is filled due to fractional averaging, what is the effect on scan time?
a. It is 25% of its original length
b. It is 75% of its original length
c. It is 125% of its original length
d. It is 175% of its original length
|
|
Definition
b. It is 75% of its original length |
|
|