Safety of the Deliver System has been Enhanced |
|
Definition
Improved mechanical and pnuematic design features
System monitors with alarms
Studies show that human error is the dominant problem in anesthetic mishap |
|
|
Term
|
Definition
PaO2< 60mmHg
Caused by either patient disease or problems with the delivery system
Ultimately machine problems result in the failure of delivery of adequate O2 to alveoli
Failure to initiate mechanical ventilation is a very common cause of hypoxia
Decreased FiO2 can be detected with oxygen analyzer |
|
|
Term
Equipment that Measures Oxygen Content in the Anesthesia Circuit |
|
Definition
Oxygen Galvanic Cell (Fuel Cell)
Polarographic Sensors
Paramagnetic Analyzer |
|
|
Term
|
Definition
Voltage produced by oxygenation of the lead electrode
Oxygen permeable membrane is used
Sensor lifetime is ~200,000% hours
CO2 can reduce life of cell
Not affected by presently used anesthetic gases |
|
|
Term
|
Definition
Proportional to O2 partial pressure |
|
|
Term
|
Definition
Used for temperature compensation
Voltage output affected by temperature and pressure |
|
|
Term
|
Definition
Uses applied current to reduce O2 to hydoxide anion
Also uses Teflon oxygen permeable membrane
The change in CURRENT is proportional to O2 in solution
Altered by circuit pressures |
|
|
Term
|
Definition
The two unpaired electrons in O2 allow attraction to a magnetic field
By applying a magnetic field to a stream of gas the pressure of the stream can be changed
O2 concentration can be calculated by changes in stream pressure |
|
|
Term
Causes of O2 Failure Upstream of the Anesthesia Machine |
|
Definition
Failure of the liquid O2 reservior
Crossed Pipelines
Closed O2 vavles
Disconnected hoses
Failure of the reserve system |
|
|
Term
Causes of O2 failure in the Anesthesia Machine or Circuit |
|
Definition
Cylinder filled with wrong gas
Cylinder Empty
Incorrect cylinder on yoke
Closed O2 valve
Failure of the proportioning system
System leak
Incompetent valves
Inadequate ventilation
Flow meter failure |
|
|
Term
|
Definition
Rarely presents a clinical problem when duration of hyperoxia < 24 hrs
May indicate a problem with the system if hyperoxia is unexpected
Easily detected with O2 analyzer
|
|
|
Term
|
Definition
Caused by patient factors or problems with the delivery system |
|
|
Term
Common Causes of Hypercarbia |
|
Definition
Apnea
Inadeqaute MV
Failure to Ventialte |
|
|
Term
Other Causes of Hypercarbia |
|
Definition
Increased dead space
Incorrectly assembled circuit system
CO2 Absorber failure
Accidential administration of CO2
Inadeqaute fresh gas flow in a non-rebreathing system |
|
|
Term
|
Definition
Unidrectional Valve Failure
Increased Areas of Bidirectional Flow |
|
|
Term
Incorrectly Assembled Circuit Sites |
|
Definition
Fresh Gas Flow
APL
Reservior Bag |
|
|
Term
|
Definition
Should not be between patient and expiratory valve |
|
|
Term
|
Definition
Should not be between patient and inspiratory valve |
|
|
Term
|
Definition
Should not be between patient and any valve |
|
|
Term
|
Definition
Bypass Valve OPen
Channeling
Exhaustion
|
|
|
Term
Inadequate Fresh Gas Flow can be caused by: |
|
Definition
Leaks cause a loss of fresh gas flow
If under minimum fresh gas flow requirements- 3-fold increase in flow rates if ventilation mode changed |
|
|
Term
|
Definition
70-100 mL/kg/min for Spontaneous ventilation |
|
|
Term
|
Definition
70-100 mL/kg/min for controlled ventilation |
|
|
Term
|
Definition
Usually a result of hyperventilation
With a bellows leak, driving gas can increase minute ventilation |
|
|
Term
Circuit Pressure and Volume Problems |
|
Definition
Occlusions
Inadequate gas in breathing system
Leaks
High Pressure in the breathing System |
|
|
Term
|
Definition
Can occur anywhere in the circuit |
|
|
Term
Inadequate gas in Breathing System |
|
Definition
Inadeqaute fresh gas flow not meeting uptake requirements
System Leaks
Excessive removal by a faulty scavenger
Excessive removal by a side stream sampler |
|
|
Term
|
Definition
Can occur at any one of multiple fittings in circuit
Most common site for a leak is Y-connector
There may be air entrainment as well as fresh gas loss |
|
|
Term
High Pressure in Breahting System |
|
Definition
Closed APL during spontaneous ventilation
Failure of ventilator to vent gas to exhaust system
Scavenger Failure
High pressure alarm should be set no higher than 50-65 cmH2O |
|
|
Term
|
Definition
Liquid Agent in Fresh Gas Piping
Incorrect vaporizer Filling
Simultaneous Use of more than one Vaporizer
Pumping Effect |
|
|
Term
Liquid Agent in Fresh Gas Piping |
|
Definition
Tilting of vaporizer
One mL of liquid can increase the concentration by 2% or more |
|
|
Term
Incorrect Vaporizer Filling |
|
Definition
A Sevo vaporizer filled with Iso set to deliver 2% will give a concentration of 3.21%
Incorrect filling of a vaporizer with Des would quickly result in lethal concentration |
|
|
Term
Simultaneous Use of More than one Vaporizer |
|
Definition
Interlock device can be defeated with incorrect vaporizer placement
Agents should be placed in order of increasing vapor pressure to minimize cross contamination |
|
|
Term
|
Definition
Almost always from failure to fill or use vaporizer |
|
|
Term
|
Definition
Present day machines use both pneumatically and electrically driven devices
The Anesthesia machine has a back up battery system which can supply 30 mins of power
The ventilator has no back up power supply, and in the event of power failure manual ventilation required |
|
|
Term
Anesthesia Machine Battery |
|
Definition
Supply lasts for 30 Mins
The battery requires 16 hrs to fully charge |
|
|
Term
|
Definition
Result of cellular metabolism and is under the control of the thermoregulation center |
|
|
Term
Normothermia is a result of |
|
Definition
Basal Metabolic Rate
Vascular tone
Muscle Activity
All factors can be affected by GA |
|
|
Term
|
Definition
Body Temperature < 36 degrees C |
|
|
Term
Consequences of Hypothermia |
|
Definition
Shivering
Increased plasma catecholamine levels
Left shift of hemoglobin
Coagulopathy
Immunosupression
Hyperglycemia
Decreased Drug Metabolism
Decreased MAC
ECG Changes |
|
|
Term
|
Definition
Can increase oxygen consumption by a factor of 5 |
|
|
Term
Increased Plasma Catecholamine Levels |
|
Definition
Vasoconstriction
Dysrhythmias
Myocardial Ischemia |
|
|
Term
ECG changes in hypothermia |
|
Definition
|
|
Term
Decreased MAC in hypothermia |
|
Definition
|
|
Term
|
Definition
Infrared rays emananting from body
Largest source of heat loss
Accounting for 40% |
|
|
Term
|
Definition
Causes vasodilation
Reduces the Shivering threshold by 2-4 Degrees C
|
|
|
Term
|
Definition
Redistribution of Heat from Core to Periphery
Steady Decline
Equilibration |
|
|
Term
|
Definition
Heat goes from core to periphery
Occurs in the 1st hour
Results in 1-2 degree C temperature decrease |
|
|
Term
|
Definition
Occurs 2-3 hours into hypothermia |
|
|
Term
|
Definition
Occurs when heat losses equal metabolic heat production |
|
|
Term
Accurate Core Temperatures |
|
Definition
PA Catheter
Distal Esophagus
Tympanic Membrane
Nasopharynx |
|
|
Term
Methods of Active Warming |
|
Definition
Heated blankets
Radiant Warmers
Forced Air Warmers |
|
|
Term
|
Definition
Shown to be most effective |
|
|
Term
|
Definition
Reported from all types of rewarming devices |
|
|
Term
|
Definition
Contraindicated in patients with Severe PVD or during hypoperfused states (aoritc cross clamping) as they will result in increased cellular activity and oxygen demand |
|
|