Shared Flashcard Set

Details

Hazards of the Delivery System
Week 8- Thursday
55
Nursing
Graduate
10/07/2015

Additional Nursing Flashcards

 


 

Cards

Term
Safety of the Deliver System has been Enhanced
Definition

Improved mechanical and pnuematic design features 

 

System monitors with alarms 

 

Studies show that human error is the dominant problem in anesthetic mishap 

Term
Hypoxia
Definition

PaO2< 60mmHg

 

Caused by either patient disease or problems with the delivery system

 

Ultimately machine problems result in the failure of delivery of adequate  O2 to alveoli

 

Failure to initiate mechanical ventilation is a very common cause of hypoxia

 

Decreased FiO2 can be detected with oxygen analyzer 

Term
Equipment that Measures Oxygen Content in the Anesthesia Circuit
Definition

Oxygen Galvanic Cell (Fuel Cell)

 

Polarographic Sensors

 

Paramagnetic Analyzer 

Term
Oxygen Galvanic Cell
Definition

Voltage produced by oxygenation of the lead electrode

 

Oxygen permeable membrane is used

 

Sensor lifetime is ~200,000% hours

 

CO2 can reduce life of cell

 

Not affected by presently used anesthetic gases 

Term
Output Voltage 
Definition
Proportional to O2 partial pressure
Term
Thermistor
Definition

Used for temperature compensation

 

Voltage output affected by temperature and pressure 

Term
Polargraphic Sensor
Definition

Uses applied current to reduce O2 to hydoxide anion 

 

Also uses Teflon oxygen permeable membrane

 

The change in CURRENT is proportional to O2 in solution 

 

Altered by circuit pressures 

Term
Paramagnetic Analyzer
Definition

The two unpaired electrons in O2 allow attraction to a magnetic field

 

By applying a magnetic field to a stream of gas the pressure of the stream can be changed

 

O2 concentration can be calculated by changes in stream pressure 

Term
Causes of O2 Failure Upstream of the Anesthesia Machine
Definition

Failure of the liquid O2 reservior

 

Crossed Pipelines

 

Closed O2 vavles

 

Disconnected hoses

 

Failure of the reserve system 

Term
Causes of O2 failure in the Anesthesia Machine or Circuit
Definition

Cylinder filled with wrong gas

 

Cylinder Empty

 

Incorrect cylinder on yoke

 

Closed O2 valve

 

Failure of the proportioning system

 

System leak

 

Incompetent valves

 

Inadequate ventilation

 

Flow meter failure 

Term
Hyperoxia
Definition

Rarely presents a clinical problem when duration of hyperoxia < 24 hrs

 

May indicate a problem with the system if hyperoxia is unexpected

 

Easily detected with O2 analyzer

 

 

Term

Hypercabia

 

 

Definition
Caused by patient factors or problems with the delivery system
Term
Common Causes of Hypercarbia
Definition

Apnea

 

Inadeqaute MV

 

Failure to Ventialte 

Term
Other Causes of Hypercarbia
Definition

Increased dead space

 

Incorrectly assembled circuit system

 

CO2 Absorber failure

 

Accidential administration of CO2

 

Inadeqaute fresh gas flow in a non-rebreathing system 

Term
Increased Dead Space
Definition

Unidrectional Valve Failure

 

Increased Areas of Bidirectional Flow

Term
Incorrectly Assembled Circuit Sites
Definition

Fresh Gas Flow

 

APL

 

Reservior Bag

Term
Fresh Gas Flow
Definition
Should not be between patient and expiratory valve
Term
APL
Definition
Should not be between patient and inspiratory valve
Term
Reservior Bag
Definition
Should not be between patient and any valve
Term
CO2 Absorber Failure
Definition

Bypass Valve OPen

 

Channeling

 

Exhaustion 

 

Term
Inadequate Fresh Gas Flow can be caused by:
Definition

Leaks cause a loss of fresh gas flow

 

If under minimum fresh gas flow requirements- 3-fold increase in flow rates if ventilation mode changed 

Term
Mapelson A
Definition
70-100 mL/kg/min for Spontaneous ventilation
Term
Mapelson D
Definition
70-100 mL/kg/min for controlled ventilation
Term
Hypocarbia
Definition

Usually a result of hyperventilation

 

With a bellows leak, driving gas can increase minute ventilation 

Term
Circuit Pressure and Volume Problems
Definition

Occlusions

 

Inadequate gas in breathing system

 

Leaks

 

High Pressure in the breathing System 

Term
Occlusions
Definition
Can occur anywhere in the circuit
Term
Inadequate gas in Breathing System
Definition

Inadeqaute fresh gas flow not meeting uptake requirements

 

System Leaks

 

Excessive removal by a faulty scavenger

 

Excessive removal by a side stream sampler 

Term
Leaks
Definition

Can occur at any one of multiple fittings in circuit

 

Most common site for a leak is Y-connector

 

There may be air entrainment as well as fresh gas loss

Term
High Pressure in Breahting System
Definition

Closed APL during spontaneous ventilation

 

Failure of ventilator to vent gas to exhaust system

 

Scavenger Failure

 

High pressure alarm should be set no higher than 50-65 cmH2O

Term
Anesthetic Overdose
Definition

Liquid Agent in Fresh Gas Piping

 

Incorrect vaporizer Filling

 

Simultaneous Use of more than one Vaporizer

 

Pumping Effect 

Term
Liquid Agent in Fresh Gas Piping
Definition

Tilting of vaporizer

 

One mL of liquid can increase the concentration by 2% or more

Term
Incorrect Vaporizer Filling
Definition

A Sevo vaporizer filled with Iso set to deliver 2% will give a concentration of 3.21%

 

Incorrect filling of a vaporizer with Des would quickly result in lethal concentration 

Term
Simultaneous Use of More than one Vaporizer
Definition

Interlock device can be defeated with incorrect vaporizer placement

 

Agents should be placed in order of increasing vapor pressure to minimize cross contamination 

Term
Anesthetic Underdose
Definition
Almost always from failure to fill or use vaporizer
Term
Electrical Failure
Definition

Present day machines use both pneumatically and electrically driven devices

 

The Anesthesia machine has a back up battery system which can supply 30 mins of power

 

The ventilator has no back up power supply, and in the event of power failure manual ventilation required 

Term
Anesthesia Machine Battery
Definition

Supply lasts for 30 Mins

 

The battery requires 16 hrs to fully charge 

Term
Heat
Definition
Result of cellular metabolism and is under the control of the thermoregulation center
Term
Normothermia is a result of
Definition

Basal Metabolic Rate

 

Vascular tone

 

Muscle Activity 

 

All factors can be affected by GA

Term
Hypothermia
Definition
Body Temperature < 36 degrees C
Term
Consequences of Hypothermia
Definition

Shivering

 

Increased plasma catecholamine levels

 

Left shift of hemoglobin

 

Coagulopathy

 

Immunosupression

 

Hyperglycemia

 

Decreased Drug Metabolism

 

Decreased MAC

 

ECG Changes 

Term
Shivering
Definition
Can increase oxygen consumption by a factor of 5
Term
Increased Plasma Catecholamine Levels
Definition

Vasoconstriction

 

Dysrhythmias

 

Myocardial Ischemia

Term
ECG changes in hypothermia
Definition

J waves

 

Ultimately V-Fib

Term
Decreased MAC in hypothermia
Definition
Prolonged PACU stay
Term
Radiation
Definition

Infrared rays emananting from body

 

Largest source of heat loss

 

Accounting for 40%

Term
One MAC of Anesthesia
Definition

Causes vasodilation 

 

Reduces the Shivering threshold by 2-4 Degrees C

 

Term
Hypothermia Phases
Definition

Redistribution of Heat from Core to Periphery

 

Steady Decline

 

Equilibration 

Term
Redistribution
Definition

Heat goes from core to periphery

 

Occurs in the 1st hour

 

Results in 1-2 degree C temperature decrease 

Term
Steady Decline
Definition
Occurs 2-3 hours into hypothermia
Term
Equilibration
Definition
Occurs when heat losses equal metabolic heat production
Term
Accurate Core Temperatures
Definition

PA Catheter

 

Distal Esophagus

 

Tympanic Membrane

 

Nasopharynx

Term
Methods of Active Warming
Definition

Heated blankets

 

Radiant Warmers

 

Forced Air Warmers 

Term
Forced Air Warmers
Definition
Shown to be most effective
Term
Burn Injuries
Definition
Reported from all types of rewarming devices
Term
External Heat Sources
Definition
Contraindicated in patients with Severe PVD or during hypoperfused states (aoritc cross clamping) as they will result in increased cellular activity and oxygen demand
Supporting users have an ad free experience!