Term
What is critical to the operation of the EM? |
|
Definition
- Generation/maintainence of the EM beam
- Without vacuum, gases would ionize and arcing would result
- Small amount of oxygen in air would cause filament to burn out
- MFP of molecule in air is 5nm vs. 6.5nm in EM
|
|
|
Term
What is the critical role for ancillary equipment? |
|
Definition
- Sputter coating
- Vacuum equipment
- FFE
|
|
|
Term
|
Definition
- Two stages
- Low vacuum followed by high vacuum
- Low vacuum (roughing vacuum) is often created by mechanical/rotary vane pump
- High vacuum is often via diffusion pump or turbomolecular pump
|
|
|
Term
|
Definition
- Gas molecules bumping into each other
- MFP is 65nm
- Roughing/mechanical pump
|
|
|
Term
|
Definition
- Gas interacts with walls of chamber
- MFP greater than column diameter
- Diffusion pump
|
|
|
Term
Describe a mechanical vacuum pump. |
|
Definition
- Physical transfer of gas from a volume
- Oil-containing
- May or may not have moving parts
|
|
|
Term
Describe an entrainment vacuum pump. |
|
Definition
- Sequesters or entrains (gas molecules still present in chamber)
- Physical or chemical means
- Non-oil containing
|
|
|
Term
Describe the rotary vane vacuum pump and its uses. |
|
Definition
- Found in nearly all EM equipment
- Works by positive displacement
- Use
- Lower the pressure to a point that a high vacuum pump can take over (works as a roughing pump)
- Remove gas from the outlet of a high vacuum pump that cannot efficiently discharge its gases (works as a backing pump)
|
|
|
Term
Describe the construction, function, and pumping characteristics of the rotary vane vacuum pump. |
|
Definition
- Construction
- Rotor with a spring-loaded vane
- Function
- As rotor turns, space between rotor and chamber wall varies
- Pumping characteristics
- 0.1Pa and 100 to 1000L/min
|
|
|
Term
Describe the vapor diffusion pump. |
|
Definition
- The most widely used vacuum pump
- Mechanical pump with no working parts
- Works by momentum transfer (=impact) rather than by diffusion
|
|
|
Term
Describe the function of the vapor diffusion pump. |
|
Definition
- Oil vapor created by heating oil
- Vapor rises
- Supersonic oil vapor knocks down gas molecules
- Gas accumulates before first annulus
- Withdrawn
- Condensation via water
- 1/2 hour warm up
- Reaches 10-4Pa at speeds of 100 to 1500 1/s
|
|
|
Term
Describe the turbomolecular pump - its advantages and characteristics. |
|
Definition
- Mechanical pump
- Provides a function similar to DP
- Advantages
- No backward diffusion of oil vapors
- No warm-up needed
- Characteristics
- Attain 10-7 to 10-8Pa at rate slower than DP
|
|
|
Term
Describe the structure and fuction of the turbomolecular pump. |
|
Definition
- Structure
- 5-15 stacks of spinning slotted rotors and stationary slotted plates (stators)
- Function
- Gas molecules hit rotor and travel to next stage
- Rotor spins at 50,000 rpm, stators encourage downward movement
- Each stage acts as a compressor stage to further concentrate the gas
- Gas is exhausted via a rotary vane pump
|
|
|
Term
|
Definition
- Entrainment via a getter surface
- Reactive gases bond to cooler getter surface
- Gettering material is often a Ti alloy with 15% Mo
- Often cooled with water or LN2
- Renewed by subliming a coat of fresh gettering metal
- Coats over previously trapped atoms
|
|
|
Term
Describe the getter pump: TSP (titanium sublimation pump) |
|
Definition
- Cylindrical chamber with sublimation filaments
- During regeneration of gettering surface, filament is heated to sublimation (requires initial vacuum of DP or TMP)
- Normal operating range for TSP is <10^-4Pa
- Used for ultrahigh vacuum gun chambers for field emission scopes
|
|
|
Term
Describe the getter pump: SIP (sputter ion pump) |
|
Definition
- Entrainment pump that ionizes gases
- Central anode of steel cylinders and cathode plates coated with gettering material (Ti)
- Incoming gas molecules are ionized by externally applied magnetic field
- Ions are unaffected
- Electrons are trapped in cylinders by magnetic field
- Ions fly directly to cathode, lose their charge and are entrained
- When ions hit, they sputter away some of the cathode material with is eventually used up
- Initial vacuum must be high (from DP or TMP)
- Used as gun chamber pumps on FE scopes
|
|
|
Term
Describe cryoadsorption pumps. |
|
Definition
- Entrainment pump
- Efficient, quiet
- Condenses some gases on cold surface and traps others in an adsorptive material
- Can reach ultrahigh vacuum of 10^-10Pa (used to pump FE gun chamber)
- Common type is two-stage helium gas refridgerator-cooled cryoadsorption pump
|
|
|
Term
Describe the first stage of the two-stage cryoadsorption pump. |
|
Definition
- Furst stage is cryocondensation
- Consists of radiation shield kept at -190C that traps water vapor and thermally shields inner stages
- May be cooled by LN2 to reduce load on helium refridgerator
|
|
|
Term
Describe the second stage of the two-stage cryoadsorption pump.
|
|
Definition
- Second stage is cryocondensation and adsorption
- At -250C; also cooled by liquid He
- Traps all gases by freezing or adsorbing them - activated charcoal, molecular sieves
- Needs to be brought to room temp once a week to pump away adsorbed and condensed gases
|
|
|
Term
Describe how a vacuum can be measured. |
|
Definition
- Need to measure vacuum in a chamber or know vacuum level for operation of control valves and pumps in automated systems
- The gauges used are indirect readings
- Measure some pressure-related property and transduce the property to electronic symbol
|
|
|
Term
Describe the operational principle of the pirani gauge. |
|
Definition
- Used for low vacuum (roughing vacuum)
- Laminar flow range or 100Pa - 10^-1Pa
- Operational principle
- Heat loss from a hot wire is transferred to surrounding gas and the resistance of wire changes
- Ammeter detects current difference which can be read as vacuum
|
|
|
Term
Describe the system of the pirani gauge. |
|
Definition
- Two resistor wires
- Reference wire (standard condition, fixed resistance) and measurement wire (under vacuum, resistance changes)
- As vacuum increases, temp of measurement wire increases
- Resistance increases and current decreases
- Resistance is detected by Wheatstone bridge circuit which applies more I to maintain R
|
|
|
Term
Describe the cold cathode gauge. |
|
Definition
- Aka Penning gauge
- Used for high vacuum
- Molecular flow range 10^0-10^-6 Pa
- Operational principle
- Ionized gas molecules create a current proportional to the vacuum
|
|
|
Term
Describe the system of the cold cathode gauge. |
|
Definition
- Voltage of several thousand volts is maintained between cathodes and anodes
- High voltage ionizes gas molecules
- Positive ions travel to negative cathode and lose their charge, causeing a current to be generated
- Current is calibrated to measure vacuum
- System is made more efficient when e- formed during initial ionization head toward + anode and ionize additional gas molecules as result of an externally applied mag. field that lengthens their time of flight
|
|
|
Term
Describe the theory and operation of the thermocouple gauge. |
|
Definition
- Used for low vacuum (roughing vacuum) - 100Pa -10^-1 Pa
- Theory and operation
- Heat lost from a hot wire is transferred to a gas
- As vacuum increases wire gets hotter
- Temperature of the wire is measured by thermocouple or voltmeter attached to thermocouple
- Temperature or current is correlated to vacuum level
|
|
|
Term
Describe the operational principle of the hot cathode gauge. |
|
Definition
- Aka Bayard-Alpert of ionization gauge
- Used for high and ultrahigh vacuum - 10^-4 to 10^-10 Pa
- Operational principle
- e- ionize gas molecules which, when collected, create a current inversely proportional to the vacuum
|
|
|
Term
Describe the system of the hot cathode gauge. |
|
Definition
- Hot filament cathode generates e- by thermionic emission
- 200V accelerates e- towards + anode (Grid=wire coil)
- e- ionize gas molecules
- Ions attracted to a central wire/ion collector
- Ions collide with collector and lose charge, create current
|
|
|
Term
Describe the two-stage rotary/DP or TMP vacuum system. |
|
Definition
- First stage: rotary rough-pump pumps column/chamber down to vacuum DP can be used
- Second stage: Valve (roughing) between rotary pump and column/chamber is closed and valve (main/plate) isolating DP is opened, backing valve opened so rotary pump backs DP by removing gas compressed in bottom stage of DP
|
|
|
Term
Describe differentially pumped systems: FE microscope, and ESEM. |
|
Definition
- May not need uniform vacuum in the system
- FE Microscope
- Gun chamber kept at ultrahigh vacuum, rest of system kept at high vacuum
- Accomplished by small diameter of EM column
- ESEM
- Specimen chamber kept at low vacuum
- series of apertures separate chamber from column so that upper parts of column above apertures can be kept at a vacuum thousands of times better than specimen chamber
|
|
|
Term
Describe the advantages and disadvantages of the two-stage rotary/oil diffusion-pumped systems. |
|
Definition
- Advantages
- Last a long time
- Can pump a lot of gas
- Disadvantages
- Back-streaming of oil vapor
- One-half hour warm-up and cool-down
|
|
|
Term
Describe the advantages and disadvantages of the turbomolecular-pumped systems. |
|
Definition
- Simplest system
- Pumps from ambient to working
- At working is backed by rotary pump
- Advantages
- Clean, fast, don't need valves of 2-stage system
- Disadvantages
- Cost, foreign material drawn into rotor, lubrication of rotor-shaft bearing
|
|
|
Term
What do lenses do? Describe light lenses and electron lenses. |
|
Definition
Lenses bend rays of light or electrons so that the rays are deflected in a predictable fasion
- Light lenses cause a slowing down of the velocity of light as it passes through the material of the lens
- Electron lenses cause a spiraling of the electrons as they pass through the bore of the lens
|
|
|
Term
Provide examples of convergent and divergent lenses. |
|
Definition
- Glass lenses of magnifying glasses, camera lenses, and LM can be convergent (positive) or divergent (negative)
- Magnetic lenses of the EM are convergent (positive)
|
|
|
Term
|
Definition
In a convergent lens, the emergent beam will converge to a focal point or common focus |
|
|
Term
|
Definition
- Distance from the lens to the focal point
- A strong lens has a short focal length
- A weak lens has a long focal length |
|
|
Term
|
Definition
- Ability to distinguish fine detail
- Minimum separation at which two objects can be seen as separate entities and not blurred together as one
- Not the same as magnification
|
|
|
Term
|
Definition
- X life size
- Maximum magnification = r.p(eye)/r.p(scope)
- Empty magnification
|
|
|
Term
Why is high contrast needed? |
|
Definition
Needed for object to stand out against background |
|
|
Term
Describe resolution limits. |
|
Definition
- Diffraction limits resolution in an aberration-free optical system
- Object size and wavelength must be of similar magnitude for diffraction to occur
- We can quantitate the magnitude of the diffraction effect using the Abbe equation .6lambda/nsinalpha
|
|
|
Term
|
Definition
- Bending of light by a barrier
- Part of wavelength that clears barrier continues in straight line (primary wavefront, undiffracted)
- Part of wavelength that hits edge of barrier goes around the corner (secondary wavefront, diffracted)
|
|
|
Term
Describe the secondary wavefront. |
|
Definition
- When wavelength hits the edge of barrier and goes around the corner - diffracted
- Is a new spherical wavefront
- The circular pattern effect is diffration
- Explained by Huygen's principle
- Each point on a wavefront may be regarded as a new source of waves
- Primary wavefront interferes with secondary
|
|
|
Term
|
Definition
- Primary wavefront interferes with secondary
- Part of advancing wavefront passes through (undiffracted or zero order)
- Edges of slit create two secondary spherical wavelet patterns close to edges of slit/pinhole
- Diffracted wave interferes with undiffracted wavefront
- Interference takes three basic forms: constructive, destructive, and partial
|
|
|
Term
Descrive diffraction fringes. |
|
Definition
- Spectral bands are produced by diffraction and constructive and desctructive interference at a slit
- Undiffracted/zero order, first order, second order, third order, and so on
- Fringes/bands caused by diffraction at small openings and sharp edges are called Fresnel fringes
- French mathematical physicist Augustin Fresnel (1788-1827)
- Produced when e- beam strikes an opaque edge
- Can be used as a focusing aid in TEM (look at small holes in a section)
|
|
|
Term
Diffraction vs. Resolution |
|
Definition
- Diffraction at a small aperture or a slit has the effect of broadening the source
- Degrades resolution
- Is the reason we don't use small aperture
|
|
|
Term
Who was Sir George Airy, and why was he important? |
|
Definition
- 19th centurary British astronomer
- Infinitely small point of light cannot be imaged as a perfect point due to an aperture
- Diffraction at the aperture gives rise to fringes that surround the image
- Pattern is called an Airy Disc
|
|
|
Term
What does the Airy disc diameter (or radius) vary with? |
|
Definition
- Aperture, wavelength, and refractive index of medium between lens and object
- 84% of incident energy found in central peak
|
|
|
Term
Who was Lord Rayleigh, and how was he important? |
|
Definition
- Radius related to resolution
- 1896 - examine resolution
- Two disc resolved when separation between centers is same as radius of first dark ring
- Need 19% intensity drop between two central peaks
- R.P. = lambda/nsinalpha
|
|
|
Term
Describe Ernst Abbe's connection to resolution. |
|
Definition
- Examine resolution in the LM
- Large aperture lens collects more diffracted light and give more information
- Introduced numerical aperture
- NA = nsina where n is refractive index of the medium and a is half the lens acceptance angle
- The Abbe equation
- R.P. = .6lambda/nsina
- Wavelength and aperture angle set limites to resolution
- R.P. is resolution of two minute pinholes in a metal film by the objective lens
|
|
|
Term
What is refraction? What is the index of refraction? |
|
Definition
- Bending of light at interface
- Index of refraction, n
- Speed of light in vacuum/speed of light in medium
- A measure of optical density
- Speed of light in glass is less than in air: glass is optically more dense
|
|
|
Term
What is the maximum resolution in the LM? |
|
Definition
- Lambda = 0.5um (green light)
- Highest NA: n=1.74 for methylene iopdide, 1.78 if S is dissolved in it, sina=0.87
- Lambda<UV is a problem
- Eyes insensitive to shorter wavelengths
- Waves shorter than UV cannot yet be bent
- Energy assocaited with radiation is harmful
|
|
|
Term
|
Definition
- Should have a resolution 100,000 times better than LM
- Why isn't it better? Because of spherical aberration
- Spherical aberration is reduced by using small aperture
- BUT, small aperture gives diffraction
- So, we compromise on aperture size
|
|
|
Term
Describe spherical aberration. |
|
Definition
- Lens surfaces are spherical: zones father from the axis have a different refractive power
- Axial rays are brought to a focus point further from the lens than medial or peripheral rays
- Ds=ks*f*a^3
- Correction: combine positive and negative lens
|
|
|
Term
Describe chromatic aberration. |
|
Definition
- Results from dispersion of lens - refractive index varies according to lambda
- Light from a n axial point will not come to a common focus
- Blue rays will be focused to a point closer to the lens
- In EM, e- have different velocities and therefore different wavelengths
- e- with greater velocity are acted on by the lens for shorter time and are deflected less: focused further along the lens axis
- Dc=kc*f*a*deltaV/V
|
|
|
Term
|
Definition
Off-axis points are imaged as conical or comet-shaped |
|
|
Term
|
Definition
- Lenses are unable to produce a single point image of an off-axis point object
- Linear images at different focus result
|
|
|
Term
Define curvature of field. |
|
Definition
- Instead of image lying in a plane, it falls upon the surface of a sphere
|
|
|
Term
|
Definition
- Magnification varies from center to periphery
- Pincusion - less magnification at the center
- Barrel - greater magnification at the center
|
|
|
Term
What are the uses of light microscopy? |
|
Definition
- Used in two main ways
- Provides magnified object image
- Permits visualization of specimen structure
- An analytical tool
- Measure brightness
- Measure length, width, area
- Count
- Determine optical properties such as refractive index or reflectance
|
|
|
Term
Describe where the energy used to form an image is in the electromagnetic spectrum. |
|
Definition
- Energy used to form image is in visible light portion of electromagnetic spectrum
- Human eyes respond to wavelengths from 730-760 nm (red/infrared) to 360-380 nm (violet)
- Greatest sensitivity in green region at about 550 nm
|
|
|
Term
What is the function of a glass lens? |
|
Definition
- Bends rays of light so that they are deflected from their original path in a predictable way
- Transparent glass causes light to slow down because the refraction between light in air (or oil) and the glass
|
|
|
Term
|
Definition
- SinI/sinR=n^2/n^1
- I=incident, R=refracted, n^1 is the refractive index of the less dense medium
|
|
|
Term
Describe the contruction/structure of a glass lens. |
|
Definition
- An optical system with two refracting surfaces
- Two main types: Converging, diverging
|
|
|
Term
Describe the ray diagram? |
|
Definition
- Used in geometrical optics to study:
- Paths followed by rays of light or e- through lenses
- Constructions used to find the relative positions and sizes of objects and their images
- By convention, rays travel from left to right
|
|
|
Term
Describe a converging lens. |
|
Definition
- Incident light rays converge after refraction
- Focal length is positive
- Called positive, convex, condensing or magnifying lens
- Always thicker at its center than the edge
- Produces a real image or a virtual image
|
|
|
Term
|
Definition
- An image that can be seen on a surface such as a screen or photographic film
- The rays intersect and physically reunite
- The image is magnified and inverted
|
|
|
Term
|
Definition
- An image that cannot be received on a surface such as a screen or film but can be converted into a real image by an optical system such as the eye, a microscope, or other converging lens system
|
|
|
Term
How is a virtual image obtained? |
|
Definition
- By placing an object between the lens and focal point
|
|
|
Term
Describe a diverging lens. |
|
Definition
- Incident light rays diverge after refraction
- Focal length is negative
- Called a negative, concave or diminishing lens
- Always thinner at center than at edges
- When used alone can only form virtual image
|
|
|
Term
Descrie the shape of converging lenses. |
|
Definition
- Meniscus
- Plano-convex
- Double-convex
|
|
|
Term
Describe the shape of diverging lenses. |
|
Definition
- Meniscus
- Plano-concave
- Double-concave
|
|
|
Term
Describe a simple microscope. |
|
Definition
- One or more lenses
- Ordinary vs. Coddington vs. Hastings triplet
- forms virtual image
- Has practical limit to magnification
- Leeuwenhoek was able to get magnifications of 200 times and resolution of 1.5um so bacteria could be seen
- Haven't been able to do better than this to date
|
|
|
Term
Describe the lenses used by the compound LM. |
|
Definition
- LM uses a series of converging and diverging lenses to produce an image
- Lenses are present in the illuminator and substage diaphragm that focus the light on teh specimen
- Lenses are present in the objectives and oculars that are used to form the image
|
|
|
Term
Describe the construction of the objective and ocular lenses in the compound LM. |
|
Definition
- Objective lens - constructed of converging and diverging elements arranged in groups
- Ocular - two converging plano-convex lenses
|
|
|
Term
What are the functions of the compound microscope lenses? |
|
Definition
- 1. The objective produces a magnified real image and the ocular produces virtual image
- 2. The real image formed by the objective serves as an object for the ocular
- An image formed by one lens can serve as the object for a second lens
|
|
|
Term
Describe the first image forming system of the compound LM. |
|
Definition
- First image forming system is the objective
- Forms intermediate image: magnified real image
- Not intercepted but is viewed magnified through a magnifier called the eyepiece
|
|
|
Term
Describe the second image forming system of the compound LM. |
|
Definition
- Eyepiece magnifies intermediate image and forms virtual image
- the lens of our eye (or a lens in a camera attachment of the microscope) converts the virtual image into a real image
|
|
|
Term
What is the thin lens formula? |
|
Definition
- Quantitation of object and image distances, magnification
- Derived from geometrical comparisons of similar triangles
- 1/f=1/a+1/b
- b/a = magnification (ratio of object and image distances)
- f=focal length
- Objective lens: real image, short f (gives real, magnified first image)
- Ocular: virtual, longer f (gives virtual image, used to produce second image)
|
|
|
Term
|
Definition
- In the substage
- Focuses light on specimen
- Abbe
- Aplanatic
- Aplanatic-achromatic
- Corrected for red and blue chromatic aberration, spherical aberration, and field curvature
|
|
|
Term
Describe the collector lens, field diaphragm, and lamp. |
|
Definition
- In the base
- Make up the iluminator
- Collector lens, field diaphragm, lamp
|
|
|
Term
Describe infinity corrected optics. |
|
Definition
- In standard objective, rays from back aperture of lens come to focus and form an image
- For infinity corrected objective, rays from back aperture are focused to infinity and do not form an image
- The tube lens (Telan lens) receives objective rays and forms real intermediate image
|
|
|
Term
Describe bright-field microscopy. |
|
Definition
- Most common viewing mode
- Ridect light passes through condeser and objective aperture and illuminates the background against which the image is seen
- Image of the specimen becomes visible by way of absorption contrast
- Specimen reduces amplitude of light passing through it (all light or certain wavelengths of light)
|
|
|
Term
What is the problem created by such objects as cells and tissues in bright-field microscopy. |
|
Definition
- Many objects absorb naturally but specimens like cells and tissues are transparent and therfore do not change the amplitude of light
- Specimen must be made absorbent with dyes (stains)
|
|
|
Term
What are the most common dyes for paraffin sections? |
|
Definition
|
|
Term
What is the most common dye for plastic sections? |
|
Definition
|
|
Term
Describe darkfield microscopy. |
|
Definition
- Aka darkground microscopy
- Direct light is prevented from passing through the objective aperture by using oblique illumination
- Specimen features appear bright against a dark background
- Undeviated light misses the objective
- No zero order contribution to image
- Image is formed from light scattered by specimen features
- Weakly diffracted light enters objective
- Usually used for small specimens such as bacteria
|
|
|
Term
Describe incident light microscopy. |
|
Definition
- Aka reflected light microscopy or epiillumination microscopy
- Illumination falls on the object from the same side as that from which the object is observed
- Microscope is equipped with epi-illuminator
- Used for flruorescence microscopy (biological specmens), geological and metallurgical specimens, examination of semiconductor devices
|
|
|
Term
Describe the mechanism of Fluorescence microscopy. |
|
Definition
- Image formed by object fluorescence
- Mechanism
- Phton absorded, e- boosted to higher energy (excitation)
- E- returns to ground state and emits quantum of light (deexcitation)
- With energy lost as heat, emitted light has less energy and longer wavelength
|
|
|
Term
What is light source is used to excite fluorescence in fluorescence microscopy? |
|
Definition
- Ultraviolet light is used to excite fluorescence and reemission is in the visible light range
- Light source and filter system required
- Mercury lamp - high pressure
- Gives 30% light at wavelength of 365 nm
|
|
|
Term
What filters are used in fluorescence microscopy? |
|
Definition
Exciter filter
- Transmits short wavelength light to exicte fluorescence in the specimen
Barrier filter
Heat filter
|
|
|
Term
What is the chromatic beam splitter used in fluorescence microscopy? |
|
Definition
- Necessary in incident light excitation
- Reflects UV light onto specimen and at the same time allows transmission of fluorescence from the specimen to the viewer
|
|
|
Term
Desecribe phase contrast microscopy. |
|
Definition
- Unstained materials can't be visualized by brightfield microscopy
- Phase contrast microscopy converts optical path differences that we can't visualize into changes in amplitude that we can see
|
|
|
Term
What are the key parts in phase contrast? |
|
Definition
- Phase plate - developed by Zernike
- Annular diaphragm
|
|
|
Term
Describe the three steps of phase contrast. |
|
Definition
-
1. Diffracted light is retarded 1/4 wavelength by specimen
-
2. Background light is advanced 1/4 wavelength by phase ring
-
This light is coming from condenser annulus
-
This light passes directly through specimen without contacting any details of structure
-
This light goes through a phase ring
-
3. Desctructive interference can occur between specimen and background light
|
|
|
Term
Why is the intensity of light through the thin part (ring) of phase plate high? |
|
Definition
- To prevent high intensity from swamping signal from thick part of plate, the ring is darkened with metallic coating
- Have positive phase contrast and negative phase contrast
- Positive phase contrast is the most common
|
|
|
Term
Describe polarized light microscopy and give examples of its use. |
|
Definition
- Image formation based on ordered molecular arrangements/uneven density
- Ordered structures appear white on a dark background
- Examples
- Mitotic spindles, actin and myosin, condensed DNA, crystals in biological specimens
- Mineral ID
- Used analytically to determine differences in optical path length that can be used to calculate refractive index
|
|
|
Term
How is the ordered structure produced in PLM? How does PLM permit double refraction of birefringence? |
|
Definition
- Produced by crystals or precisely aligned molecules - more dense in one direction that the other
|
|
|
Term
What to birefringent materials have the power to do? |
|
Definition
- Have the power to absorb one set of the rays produced
- The orginary ray is absorbed and the extraordinary is transmitted
- We make use of this property in polarizers, analyzers, and Nicol/Wollaston prisms
|
|
|
Term
What are the parts of the PLM and their functions? |
|
Definition
- PLM is a compound LM fitted with a polarizer, analyzer, and compensator
- Compensator is used for quantitative measurement of birefringence
- Polarizers and analyzers are often nicol prisms
- Light rays emerge vibrating in a single plane
- the polarizer lets light rays vibrating in a certain plane pass through which in turn excites light rays in the crystal which can be absorbed or transmitted by the analyzer
- The crystal appears light on a dark background
|
|
|
Term
Describe differential interference contrast and its parts. |
|
Definition
- Aka Nomarski differential interference contrast
- Produces amplitude contrast from a transparent specimen
- The image produced has a distinctive relief-like, shadowed appearance that appears 3D
- Parts
- Polarizer, analyzer, Wollaston (Nomarski/DIC prisms) split and recombine light beam
|
|
|
Term
Describe the process of differential interference contrast. |
|
Definition
- Light from the light source is polarized by a polarizing filter in the substage (polarized light)
- Light emerging from the polarizer is split into two components polarized at right angle to each other
- Specimen is sampled by pairs of closely spaced rays generated by lower DIC prism
- When ray pairs traverse region with a refractive index or thickness difference, an optical path difference is introduced between the two rays
- Optical path difference is translated into amplitude difference (path differences of lambda/10 to full wavelegth permit interference, thus imaging transparent detail
- Light beams recombined by upper Nomarski prism
- Analyser above prism passes certain amount of light depending on its orientation
- Out of phase light interferes at image plane to create image
|
|
|
Term
Describe the dissecting microscope. |
|
Definition
- Aka stereomicroscope
- Used for erect, magnified image that shows depth - has considerable depth of field
- Binocular head with common objective lens
- Much of the light travels through objective at angle so objective must be of high quality
- Disadvantage is that NA is limited by double beam path to about 0.1
|
|
|
Term
Describe Kohler illumination. |
|
Definition
- Introduced by August Koehler in 1893
- Added collector lens for lamp
- Used collector lens to focus lamp image on the fron aperture of the condenser
- the field stop was then focused on the specimen with the condenser control
- Provided bright, even illumination and parallel unfocused light through the objective plane
- Gives wide cone o0f light for optimum resolution and reduces effect of dust from condenser
|
|
|
Term
What are the steps of Kohler illumination? |
|
Definition
- 1. Focus lamp on front aperture of condenser
- 2. Focus the specimen
- 3. Focus condenser to see field stop diaphragm (kohler illumination)
- Adjust condenser diaphragm using eyepiece telescope
|
|
|
Term
List types of objectives. |
|
Definition
- Plan
- Achromat
- Chromatic aberration limited to two wavelengths: blue and red
- Apochromat
- Chromatic aberration is minimized for three wavelengths (blue, green, red)
|
|
|
Term
Describe the objective characteristics. |
|
Definition
- Are indicated on lens barrel
- Indicated coverslip thickness (this is the standard number 1.5 coverslip)
- Working distance, magnification, application, numerical aperture/immersion medium
- Microscope slides should be 1.1mm thick
|
|
|