Term
Equation for 5% transmission using a number of HVL's, and solve. |
|
Definition
1/2n = 0.05
where n is the number of HVL's.
2n= 1/0.05
2n= 20
nlog2=log20, n=4.32HVL |
|
|
Term
Ballpark idea of required thickneses to provide 5% transmission Co60 6MV 10MV 25MV (in lead) |
|
Definition
Co60-5cm 6MV-6.5cm 10MV-7cm 25MV-7cm |
|
|
Term
Cerrobend melting point vs lead? |
|
Definition
|
|
Term
Density for Cerrobend vs lead? |
|
Definition
|
|
Term
When you half-beam block what happens to the isodose lines near the surface? |
|
Definition
Throws a hotspot to the outer edge of the field. |
|
|
Term
How many pairs do MLC come in and what are their sizes? |
|
Definition
26, 40, or 60 leaf pairs.
Come in size of 0.5m or 1.0cm |
|
|
Term
What is the transmission of MLC leaves? In between leaves? |
|
Definition
<1% transmission through leaves
<1-2% between leaves
|
|
|
Term
What is the transmission through the jaws of the machine? |
|
Definition
|
|
Term
Is Penumbra larger or smaller with MLC vs blocks? |
|
Definition
Penumbra with MLC is larger. |
|
|
Term
By what photon interaction contributes to skin dose? |
|
Definition
Anything put into the beam will produce Compton electrons. |
|
|
Term
What instruments do we use to measure Dose Distributions in the buildup region? |
|
Definition
Extrapolation chambers are best but not widely available.
Many use fixed seperation plane-parallel ionization chambers.
TLDs are another choice (LiF) |
|
|
Term
|
Definition
|
|
Term
What are electron filters? |
|
Definition
When placed in photon beam, they are designed to attenuate secondary e-'s . Made of medium z material (copper, tin, lead) |
|
|
Term
What is the purpose of electron filters and for what energy? |
|
Definition
They help preserve skin sparing and are useful in large fields on 60Co machines. |
|
|
Term
|
Definition
Gap seperation for adjacent fields
S1=½*L1*[d/SSD1]
S2=½*L2*[d/SSD2] |
|
|
Term
Total seperation for adjacent fields |
|
Definition
Total separation =
(½L1 . 1/SSD1) + (½L2 . 1/SSD2) |
|
|
Term
How do we eliminate the "three-field overlap"
(when L1 and L2 are different lengths) using AP/PA fields? |
|
Definition
ΔS' = ΔS x d'-d/d
where d'=depth of cord from surface.
d= depth of Midline
and ΔS= S1-S2.
New gap required = S1+S2+ΔS' |
|
|
Term
Craniospinal collimator and couch kick equation.
|
|
Definition
1. Careful not to exit through mouth!
Collimator = arctan(½(Lspinefld)*1/SSD)
Couch kick= arctan(½(LWBlength)*1/SAD) |
|
|