Ability to image accurately |
|
|
Term
|
Definition
The ability to resolve two separate images along (parallel to) the main axis of a sound beam
- Units: distance
- Determined by: spatial pulse length
- Shorter pulses improve resolution
- Reduce ringing (fewer cycles/pulse)
- Increase frequency/decrease wavelength
- Shorter pulse duration also improves resolution
- Not adjustable -> same at all depths
- Range: 0.1-1.0 mm -> the smaller, the better
|
|
|
Term
|
Definition
Synonyms for axial resolution
- Longitudinal
- Axial
- Range
- Radial
- Depth
|
|
|
Term
Formulas for Axial Resolution |
|
Definition
= SPL (mm) ÷ 2
= (wavelength (mm) * #cycles/pulse) ÷ 2
In soft tissue: = (0.77 * #cycles/pulse) ÷ frequency (MHz) |
|
|
Term
|
Definition
The ability to resolve two separate images that are side by side (perpendicular) to the main axis of a sound beam
- Units: distance
- Determined by: sound beam diameter
- Narrower beams improve resolution
- Adjustable depth by changing focus
|
|
|
Term
|
Definition
Synonyms for lateral resolution
- Lateral
- Angular
- Transverse
- Azimuthal
|
|
|
Term
|
Definition
Concentrates the sound energy into a narrower beam
Alters the beam by:
- Narrowing beam diameter in near field and focal zone
- Decreasing focal depth (shallower focus)
- Shrinking focal zone (the greater the extent, the smaller the zone)
- Increasing beam diameter in far zone
|
|
|
Term
|
Definition
- Fixed aka conventional/mechanical
- Focal depth and extent of focusing cannot be changed
- Includes lens and curved active elements
- Single crystal transducers
- Poorest lateral resolution
- Phased array aka adjustable/multiple focusing
- Adjustable by electronics
- Generally better lateral resolution
|
|
|
Term
Three Methods of Focusing |
|
Definition
- External -> with a lens
- Internal -> with a curved active element
- Phased array -> with electronics of ultrasound system
|
|
|