function of AEC devices
Definition
eliminate need for radiographer to set exposure time
Term
Phototimer
Definition
beam passes through patient, tabletop, IR, and THEN photodetector
calibrated by setting capacitor discharge at point that producies satifactory density
Term
components of a phototimer
Definition
fluorescent screen
photomultiplier tube
capacitor
thyratron
Term
fluorescent screen
Definition
absorbs xrays and produces light photons
Term
photomultiplier tube
Definition
produces electrons form the light photons received from fluorescent screen
Term
capacitor
Definition
discharges when reaches present value, triggering thyratron and opening relay to terminate exposure
Term
thyratron
Definition
regulates exposure automatically
Term
ionization chamber
Definition
beam passes through patient and table top before striking the ionization chamber
Term
components of ionization chamber
Definition
flat, parallel plate ionization chamber
made radiolucent
Term
ionization within the chamber creates a
Definition
charge
Term
exposure is terminated when
Definition
appropriate charge has been reached
Term
most creitical element in using AECs
Definition
exact position of ion chambers must be known to radiographer
Term
backup timer
Definition
time the electronic timer is set in case AEC fails to terminate exposure
cannot exceed tube limit and should be set 150% of anticipated mAs
Term
regulations require that AEC has ___ mAs and ___ mAs safety override
Definition
600, 2000