Term
What is the ICRU clinical Dose Accuracy? |
|
Definition
|
|
Term
What are the factors contribute to both random and systematic deviations in dose delivery? |
|
Definition
Daily patient setup
Target delineation
Dose calculation are the main factors |
|
|
Term
What is the main limitation of the Manual dose calculation? |
|
Definition
Manual calculations are based on machine data, which are typically gathered with a flat, homogeneous water phantom. |
|
|
Term
What is the advangage of image-based
treatment planning System? |
|
Definition
The Image-based treatment planning has allowed incorporation of patientspecific internal heterogeneity information into the calculation of dose.
|
|
|
Term
The normalization depth for photon and electron dosimetry |
|
Definition
For photons, d0 = 10 cm is recommended, but not required. For each photon beam, d0 is independent of field size and shall be greater than or equal to the maximum dm.
For electrons, d0 is taken to be the dm along the central axis for the same field incident on a water phantom at the same SSD. It is field-size dependent.
|
|
|
Term
Does this protocol require D’0 =1 cGy/MU ?
|
|
Definition
|
|
Term
What are the advantages of choosing a normalization depth of 10 cm? |
|
Definition
Choosing d0 = 10 cm for photon beams eliminates the uncertainty associated with converting the calibrated output to the dose rate at other depths, particularly at dm
Different machines of the same energy will be matched at a more clinically relevant depth, which may decrease the differences in programmed MU when moving patients from one machine to another.
Some of the field size dosimetric quantities vary less significantly at depths greater than dm, making dosimetric measurements less susceptible to setup error.
Some treatment-planning systems require measured output factors at a depth of 10 cm, thus requiring users to measure these data anyway. Choosing a normalization depth of 10 cm eliminates the duplication of effort, either at time of commissioning or during annual inspections.
|
|
|
Term
What is the requirements of the d0 other than 10 cm? |
|
Definition
If another depth is chosen for d0, at a minimum this depth shall be greater than or equal to the maximum dm depth, determined from percent depth dose measurements for the smallest field size and greatest SSD used clinically.
|
|
|
Term
Photon calculations using tissue phantom ratio (TPR). |
|
Definition
|
|
Term
|
Definition
|
|