Term
____________ or __________ produces replications of moving anatomy across the image in the phase encoding direction.
225 |
|
Definition
Phase mismapping, ghosting |
|
|
Term
As ___________ only occurs during the phase axis, the direction of phase encoding can be changed, so that the artefact does not interfere with the area of interest.
227 |
|
Definition
|
|
Term
______________ nulls signal from specified areas. Placing " " volumes over the area producing artefact nullifies signal and reduces the artefact.
228 |
|
Definition
|
|
Term
When imaging the chest and abdomen, a method known as ______________ or ____________ can greatly reduce ghosting from respiration.
230 |
|
Definition
respiratory compensation, respiratory ordered phase encoding (ROPE) |
|
|
Term
The only penalty for using the respiratory compensation method is that the number of slices available for a given TR may be slightly reduced/increased.
230 |
|
Definition
|
|
Term
Systems also use a method known as _______________ or __________ that times the excitation RF with a certain phase of respiration.
231 |
|
Definition
respiratory gating, triggering |
|
|
Term
___________ is a very general term used to describe a technique of reducing phase mismapping from the periodic motion caused by respiration, cardiac and pulsatile flow motion.
233 |
|
Definition
|
|
Term
_________________ gating uses electrodes and lead wires that are attached to the patient's chest to produce an ECG. This is used to determine the timing of the application of each excitation pulse.
233 |
|
Definition
Electrocardiogram (ECG, EKG) |
|
|
Term
____________ gating uses a light sensor attached to the patient's finger to detect the pulsation of blood through the capillaries.
233 |
|
Definition
|
|
Term
__________________ reduces ghosting caused by flowing nuclei moving along gradients.
233 |
|
Definition
|
|
Term
___________ or ___________ produces an image where anatomy that exists outside the FOV is folded onto the top of anatomy inside the FOV.
234 |
|
Definition
|
|
Term
Aliasing along the frequency encoding axis is known as _____________. This is caused by undersampling the frequencies that are present in the echo.
235 |
|
Definition
|
|
Term
Aliasing along the phase axis of the image is known as _____________. This is caused by undersampling along the phase axis. |
|
Definition
|
|
Term
A ____________ is used to filter out frequencies that occur outside the selected FOV. Signal originating from outside the FOV along the frequency axis is no longer mismapped as it is filtered out.
239 |
|
Definition
|
|
Term
Anti-aliasing along the phase axis is termed___________, ______________, or _____________, which oversamples along the phase encoding axis by increasing the number of phase encodings performed.
240 |
|
Definition
no phase wrap, phase over-sampling, anti-foldover |
|
|
Term
______________ produces a dark edge at the interface between fat and water. It occurs along the frequency encoding axis only.
243 |
|
Definition
|
|
Term
Chemical shift can be limited by scanning at lower/higher field strengths and by keeping the FOV to a minimum/maximum.
244 |
|
Definition
|
|
Term
To reduce chemical shift artefact always use the narrowest/widest bandwidth in keeping with good SNR and the smallest/largest FOV possible.
244 |
|
Definition
|
|
Term
At _______ or less, chemical shift artefact is insignificant and usually does not need to be compensated for.
244 |
|
Definition
|
|
Term
When fat and water are in phase and their signals add constructively, and when they are out of phase their signals cancel each other out. This cancellation effect is know as _________________ or _______________.
244 |
|
Definition
out of phase artefact, chemical misregistration |
|
|
Term
To remedy chemical misregistration, select a __________ that matches the periodicity of fat and water at your field strength.
247 |
|
Definition
|
|
Term
At 1.5 T, for example, selecting a TE that is a multiple of _______ ms reduces chemical misregistration artefact, while at 0.5 T the periodicity of fat and water is ______ ms.
247 |
|
Definition
|
|
Term
___________ artefact produces a banding artefact at the interfaces of high and low signal.
249 |
|
Definition
|
|
Term
_______________ artefact results from under-sampling of data (too few K space lines filled) so that interfaces of high and low signal are incorrectly represented on the image.
249 |
|
Definition
|
|
Term
To remedy truncation artefact, under/over-sampling of data must be avoided. To do so, increase the number of phase encoding steps.
250 |
|
Definition
|
|
Term
_________________ artefact produces distortion of the image together with large signal voids.
250 |
|
Definition
|
|
Term
To remedy magnetic susceptibility artefact, the technologist should:
251 |
|
Definition
remove all metal objects
use spin echo sequences instead of gradient echo
decrease the TE |
|
|
Term
___________ and ____________ artefacts appear as adjacent slices in an acquisition having different image contrasts.
252 |
|
Definition
Cross-excitation and cross-talk |
|
|
Term
Adjacent slices receive energy from the RF excitation pulse of their neighbors. This energy pushes the NMV of the nuclei towards the transverse plane, so that they may become saturated when they themselves are excited. This effect is called _______________ and affects image contrast.
253 |
|
Definition
|
|
Term
Nuclei lose their energy due to spin lattice relaxation and may dissipate this energy to nuclei in neighboring slices. This is called ________________.
253 |
|
Definition
|
|
Term
___________ can never be eliminated as it is caused by the natural dissipation of energy by the nuclei.
253 |
|
Definition
|
|
Term
________ artefact appears as a dense line on the image at a specific point.
255 |
|
Definition
|
|
Term
__________ artefact is caused by a leak in teh RF shielding of the room.
256 |
|
Definition
|
|
Term
__________ artefact produces a difference in signal intensity across the image.
256 |
|
Definition
|
|
Term
The main cause of ___________ artefact is uneven excitation of nuclei within the patient due to RF pulses applied at flip angles other than 90º and 180º.
256 |
|
Definition
|
|
Term
To remedy ___________ artefact, always ensure that the coil is loaded correctly and that the patient is not touching the coil at any point.
256 |
|
Definition
|
|
Term
_________ artefact is shown as a black and white banding artefact on the edge of the FOV. It is always seen in gradient echo imaging.
256 |
|
Definition
|
|
Term
The cause of ___________ artefact is a combination of wrap and field inhomogeneity in gradient echo sequences.
256 |
|
Definition
|
|
Term
To remedy ___________ artefact, use spin echo sequences or ensure the patient keeps their arms within the FOV.
257 |
|
Definition
|
|
Term
____________ is seen in tissues that contain collagen (such as tendons) as high signal intensity.
257 |
|
Definition
|
|
Term
___________ artefact is caused when structures that contain collagen lie at an angle of 55° to the main field.
257 |
|
Definition
|
|
Term
To remedy ___________ artefact, alter the angle of the structure or change the TE.
258 |
|
Definition
|
|